5 Determination mode

797 VA Computrace – Software

132

i.delta; P.delta; delta; Q.delta(Only with modes CVS

and CPVS)

Difference of two successive mean values of the evaluation
quantities.

Comments

Display of comments if any type of error appeared in the
sweep (e.g.

Ovl. in scan

= Overload during sweep;

Ovl. in

CDE

= Overload during cleaning or deposition;

Ovl. in cond.

cycles

= Overload during conditioning cycles;

No peak

found

= no peak found for defined substance;

Not used

=

peak is not used for calculation).

Note: For CVS and CPVS, two additional evaluation values are dis-
played: V(mL) and Q/Q(0).

Calibration data

Substance

Substance name (see

Substance

on the

Substances

tab,

section 5.2).

Calibr.

Calibration

techniques

std.add.

,

rec.cc.

or

smp.cc.

(see

Cali-

bration

on the

Specifications

tab, section 5.2).

For electroplating bath analytics:

Calibration

techniques

dt.rec.cc.

,

dt.cc

,

mlat

,

lat rec.rc

,

sam.rc

(see Calibration

techniques with CVS and CPVS, section 6.2).

Y.reg/offset

Evaluation quantity for the sample calculated from the
standard addition curve (for standard addition) or intercept
of the calibration curve (for calibration curves).

Slope

Slope of the standard addition or calibration curve.

Nonlin

Nonlinear factor for nonlinear calibration curves (only dis-
played, if "Nonlinear Regression" was chosen for parame-
ter

Regression technique

).

Corr.Coeff

Corrrelation Coefficient for linear calibration curves (only
displayed, if "Linear Regression" was chosen for parameter

Regression technique

).

Mean deviat.

Calculated mean deviation of the measured values about
the standard addition or calibration curve.

Solutions

No.

Number of

Dosing Device

used for addition or predose.

Content

Remarks regarding the solution (see

Content

on the

DOSI-

NOS