Dt record calibration curve – Metrohm 797 VA Computrace User Manual
Page 186

6 Electroplating Bath VA
797 VA Computrace – Software
174
DT Record calibration curve
“DT Record calibration curve” is used to record calibration curves
for the dilution titration calibration technique (with the CVS and
CPVS mode). Dilution titration
Calibration
technique is mainly used
to analyze Suppressors in electroplating baths. The calibration
curves recorded with this
Calibration
technique are used to de-
termine Suppressor concentrations with the
Calibration
technique
“DT Suppressors with calibration curve”.
Achtung:
A determined calibration curve can only be adopted
automatically if the checkbox
Save calibration curves additionally
without date and time
on the
General
tab of the
GENERAL SET-
TINGS
window is checked.
Determination tab with Calibration technique “DT Record
calibration curve”
The
Determination
tab of the
EDIT WORKING METHOD PARAME-
TERS
window contains general specifications for performing the
determination. The parameters displayed depend on the selected
Calibration
technique and measurement
Technique
. With “DT Re-
cord calibration curve” following parameters have to be
defined:
Sample identifier [ 32 characters ; "sample" ]
Identification for sample.
Note: The
Sample identifier
is used for the file na-
me. To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, this
Sample identifier
must match with
the name of the calibration file defined for the pa-
rameter
Calibration curve
on the
Determination
tab with calibration "DT Suppressors with calibra-
tion curve".
Example for
Sample identifier
:
Calibration Suppressor
Note: To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, the path of the calibration file defined for
the parameter
Calibration curve
on the
Substances
tab with
Calibration
technique "DT Suppressors
with calibration curve”, must match with the path
defined for the parameter
Data folder
(for the cur-
rently logged-in user) on the tab
User Directories
of the window
USER RIGHTS
.
Note: A determined calibration curve is only
adopted automatically if the checkbox
Save cali-