6 Electroplating Bath VA

797 VA Computrace – Software

174

DT Record calibration curve

“DT Record calibration curve” is used to record calibration curves
for the dilution titration calibration technique (with the CVS and
CPVS mode). Dilution titration

Calibration

technique is mainly used

to analyze Suppressors in electroplating baths. The calibration
curves recorded with this

Calibration

technique are used to de-

termine Suppressor concentrations with the

Calibration

technique

“DT Suppressors with calibration curve”.

Achtung:

A determined calibration curve can only be adopted

automatically if the checkbox

Save calibration curves additionally

without date and time

on the

General

tab of the

GENERAL SET-

TINGS

window is checked.

Determination tab with Calibration technique “DT Record
calibration curve”

The

Determination

tab of the

EDIT WORKING METHOD PARAME-

TERS

window contains general specifications for performing the

determination. The parameters displayed depend on the selected

Calibration

technique and measurement

Technique

. With “DT Re-

cord calibration curve” following parameters have to be

defined:

Sample identifier [ 32 characters ; "sample" ]

Identification for sample.

Note: The

Sample identifier

is used for the file na-

me. To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, this

Sample identifier

must match with

the name of the calibration file defined for the pa-
rameter

Calibration curve

on the

Determination

tab with calibration "DT Suppressors with calibra-
tion curve".

Example for

Sample identifier

:

Calibration Suppressor

Note: To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, the path of the calibration file defined for
the parameter

Calibration curve

on the

Substances

tab with

Calibration

technique "DT Suppressors

with calibration curve”, must match with the path
defined for the parameter

Data folder

(for the cur-

rently logged-in user) on the tab

User Directories

of the window

USER RIGHTS

.

Note: A determined calibration curve is only
adopted automatically if the checkbox

Save cali-