8.6 Analyze Electroplating Bath Solutions

797 VA Computrace – Software

239

Note: To make sure that always the latest calibration file is ta-
ken for the calculation, the path of the calibration file defined
for the parameter

Calibration curve

on the

Determination

tab (of

the window

EDIT WORKING METHOD PARAMETERS

with

cali-

bration

"DT Suppressors with calibration curve", must match

with the path defined for the parameter

Data folder

(for the cur-

rently logged-in user) on the tab

User Directories

of the window

USER RIGHTS

.

Sample table
In the Sample table, list (according to the arrangement on the sam-
ple rack) for every Suppressor standard solution a method with

Calibration

technique "DT Record calibration curve" and for every

sample a method with

Calibration

technique "DT Suppressors with

calibration curve":

Start determination.

Suppressor Analysis with 838 Advanced Sample Processor and RC

Following installations and settings are recommended for the au-
tomated Suppressor determination with the 838 Advanced
Sample Processor

and the "response curve technique":

Note: The "response curve technique" is used for Suppressor de-
termination in electroplating baths, if the “dilution titration techni-
que" is inapplicable.

Instruments
Install the 838 Advanced Sample Processor, two 800 Dosi-
nos
and a 732 Relay Box with two 823 Membrane Pump
Units
(see Hardware Manual 797 and 838 Instructions for Use).

Dosinos:

Dosino 1:

50 mL Exchange Unit

Electrolyte

solution