Metrohm 797 VA Computrace User Manual
Page 68
3 General settings for exploratory and determination mode
797 VA Computrace – Software
56
Sweep parameters:
Hydrodynamic (measurement) [ on, off ; on ]
Enable/disable stirring of the RDE/SSE during the sweep.
Start potential (V) [ -5...+5 V ; 1.625 V ]
First turning point for the potential sweep.
Note: The Start potential needs to be between the
First ver-
tex potential
and the
Second vertex potential
.
First vertex potential (V) [ -5...+5 V ; -0.175 V ]
First turning point for the potential sweep.
Second vertex potential (V) [ -5...+5 V ; 1.625 V ]
Second turning point for the potential sweep (Can be differ-
ent from the
Start potential
).
Voltage step (V)
[ > 0 V ; 0.006 V ]
Voltage step for direct voltage ramp.
Sweep rate (V/s)
[ > 0 V/s ; 0.1 V/s ]
Ramp slope =
Voltage step / Voltage step time
.
No. of sweeps
[ > 0 ; 3 ]
Number of cyclic sweeps to be performed.
Save last ... sweeps
[ > 0 ; 2 ]
Number of cycles to be saved.
Note: Total number of saved sweeps is number of
Save last
sweeps
multiplied with
No. of replications
.
Notes:
•
The CVS measurement mode is mainly used with RDE elec-
trodes. HMDE is possible.
•
The following condition applies to the definition of
Voltage
step
and
Sweep rate
:
Voltage step time = Voltage step / Sweep rate > 270 μs
•
The measurement time
t (i)
is defined as follows:
Voltage step time ≥ 80 ms
→
t (i)
=
20/16.67 ms
(power fre-
quency 50/60 Hz)
Voltage step time < 80 ms
→
t (i)
=
Voltage step time / 4
(Volt-
age step time = Voltage step / Sweep rate)