Page 68

background image

3 General settings for exploratory and determination mode

797 VA Computrace – Software

56

Sweep parameters:

Hydrodynamic (measurement) [ on, off ; on ]

Enable/disable stirring of the RDE/SSE during the sweep.

Start potential (V) [ -5...+5 V ; 1.625 V ]

First turning point for the potential sweep.

Note: The Start potential needs to be between the

First ver-

tex potential

and the

Second vertex potential

.

First vertex potential (V) [ -5...+5 V ; -0.175 V ]

First turning point for the potential sweep.

Second vertex potential (V) [ -5...+5 V ; 1.625 V ]

Second turning point for the potential sweep (Can be differ-
ent from the

Start potential

).

Voltage step (V)

[ > 0 V ; 0.006 V ]

Voltage step for direct voltage ramp.

Sweep rate (V/s)

[ > 0 V/s ; 0.1 V/s ]

Ramp slope =

Voltage step / Voltage step time

.

No. of sweeps

[ > 0 ; 3 ]

Number of cyclic sweeps to be performed.

Save last ... sweeps

[ > 0 ; 2 ]

Number of cycles to be saved.

Note: Total number of saved sweeps is number of

Save last

sweeps

multiplied with

No. of replications

.

Notes:

The CVS measurement mode is mainly used with RDE elec-
trodes. HMDE is possible.

The following condition applies to the definition of

Voltage

step

and

Sweep rate

:

Voltage step time = Voltage step / Sweep rate > 270 μs

The measurement time

t (i)

is defined as follows:

Voltage step time 80 ms

t (i)

=

20/16.67 ms

(power fre-

quency 50/60 Hz)

Voltage step time < 80 ms

t (i)

=

Voltage step time / 4

(Volt-

age step time = Voltage step / Sweep rate)