8 How to ...?
797 VA Computrace – Software
246
gle or mixed standard addition solution, its concentration and
its volume is defined.
10. If the standard addition should be done with variable addition
volumes, click the button in the
Volume
column to open
the
EDIT VARIED ADDITION
window, enter the variable addi-
tion volumes in the
Addition
fields, and close this window by
clicking
.
11. Close the
EDIT WORKING METHOD PARAMETERS
window by
clicking
.
12. Place the sample solution in the measuring vessel at the 797
VA Computrace stand.
13. Click on
or
MAIN WINDOW
/ Window / Monitor
to open
the
MONITOR
window.
14. Start the measurement by clicking the
icon in the
MAIN
WINDOW
or the
button in the
MONITORING
win-
dow.
15. Enter the
Sample ID
(used as part of the determination file na-
me) in the
PLACE SAMPLE
window and click
.
16. Each time a standard addition is required in the
MANUAL AD-
DITION
window, add the standard addition solution using a pi-
pette.
Use manual standard addition with solution exchange
In the manual standard addition with solution exchange, a new
sample solution is used for every standard addition. Proceed as fol-
lows:
1. Click on
or
MAIN WINDOW
/ Mode / Determination
.
2. Click on
or
MAIN WINDOW
/ Window / Working method
specification
to open the
WORKING METHOD SPECIFICATIONS
window.
3. Load the desired method into the
WORKING METHOD SPECIFI-
CATIONS
window (see How to Load a method, section 8.4).
4. Select
Standard addition
in the
Calibration
field of the
WORK-
ING METHOD SPECIFICATIONS
window.
5. Select
Manual
in the
Addition
field of the
WORKING METHOD
SPECIFICATIONS
window.
6. Select
Batch with solution exchange
in the
Technique
field of
the
WORKING METHOD SPECIFICATIONS
window (is not select-
able with the CVS and CPVS mode).
7. Click
to open the
EDIT WORKING METHOD PA-
RAMETERS
window.