Stand-by potential – Metrohm 797 VA Computrace User Manual
Page 75

3.4 General operation sequence
797 VA Computrace – Software
63
Cleaning potential (V)
[ -5...+5 V ; -0.1 V ]
Voltage applied to the electrodes during the
Cleaning time
.
Cleaning time (s)
[ 0...80600 s ; 0 s ]
Time during which the
Cleaning
potential
is applied to the
electrodes.
Deposition potential (V)
[ -5...+5 V ; -0.9 V ]
Voltage applied to the electrodes during the
Deposition time
.
Deposition time (s)
[ 0...80600 s ; 60 s ]
Time during which the
Deposition
potential
is applied to the
electrodes.
Equilibration time (s) [ 0...80600 s ; 5 s ]
Waiting time before starting the sweep; with start potential
applied to the electrodes. If the checkbox
Hydrodynamic
(measurement)
is not checked, the stirrer is switched off dur-
ing that time.
Achtung: In the CVS and CPVS mode, no deposition is pos-
sible, and an own Equilibration potential can be defined (see
Pretreatment with CVS and CPVS). With CPVS, it is part of the
sweep.
Stand-by potential
The stand-by potential can be applied to the electrodes at the
end of the measurement. It remains in force until it is switched off
manually in the
COMPUTRACE CONTROL
window or until a new
voltage is applied to the electrodes in the next measurement.
Cell off after measurement
[ on, off ; on ]
Enable/disable the switching off of the voltage applied to the
electrodes after measurement.
Stand-by potential (V)
[ -5...+5 V ; -0.1 V ]
Voltage to be applied to the electrodes after measurement if
the
Cell off after measurement
box
is not activated.
Note: If
Cell off after measurement
is not checked, and you
want to remove any electrode cable, the
Stand-by potential
must be turned off by clicking the
button on the
COMPUTRACE CONTROL
window.