3.4 General operation sequence

797 VA Computrace – Software

63

Cleaning potential (V)

[ -5...+5 V ; -0.1 V ]

Voltage applied to the electrodes during the

Cleaning time

.

Cleaning time (s)

[ 0...80600 s ; 0 s ]

Time during which the

Cleaning

potential

is applied to the

electrodes.

Deposition potential (V)

[ -5...+5 V ; -0.9 V ]

Voltage applied to the electrodes during the

Deposition time

.

Deposition time (s)

[ 0...80600 s ; 60 s ]

Time during which the

Deposition

potential

is applied to the

electrodes.

Equilibration time (s) [ 0...80600 s ; 5 s ]

Waiting time before starting the sweep; with start potential
applied to the electrodes. If the checkbox

Hydrodynamic

(measurement)

is not checked, the stirrer is switched off dur-

ing that time.

Achtung: In the CVS and CPVS mode, no deposition is pos-
sible, and an own Equilibration potential can be defined (see
Pretreatment with CVS and CPVS). With CPVS, it is part of the
sweep.

Stand-by potential

The stand-by potential can be applied to the electrodes at the
end of the measurement. It remains in force until it is switched off
manually in the

COMPUTRACE CONTROL

window or until a new

voltage is applied to the electrodes in the next measurement.

Cell off after measurement

[ on, off ; on ]

Enable/disable the switching off of the voltage applied to the
electrodes after measurement.

Stand-by potential (V)

[ -5...+5 V ; -0.1 V ]

Voltage to be applied to the electrodes after measurement if
the

Cell off after measurement

box

is not activated.

Note: If

Cell off after measurement

is not checked, and you

want to remove any electrode cable, the

Stand-by potential

must be turned off by clicking the

button on the

COMPUTRACE CONTROL

window.