6 Electroplating Bath VA

797 VA Computrace – Software

194

Production bath solution

The “Production bath solution” (or “Plating bath solution”) is the
sample solution to be analyzed. It is used in the Electroplating Bath
analysis with the modes CVS and CPVS.

Addition ratio

The “Addition ratio” is a certain Q/Q(0) ratio which defines the
number of additions with the two “dilution titration technique”

Ca-

libration

techniques “DT Record calibration curve” and “DT Supp-

ressors with calibration curve” (with the modes CVS and CPVS). It
should normally be around 0.3 and can be defined on the

Substan-

ces

tab of the

EDIT WORKING METHOD PARAMETERS

window.

The “Addition ratio” defines the criteria for the endpoint. After
reaching this point, the standard addition is stopped and the “Cali-
bration factor” is calculated.

Note: The following condition must be fulfilled: „Begin of evalua-
tion“ > “Evaluation ratio” > „Addition ratio“.

Evaluation ratio

The “Evaluation ratio” is a certain Q/Q(0) ratio and defines the
point of evaluation for the calculation of the “Calibration factor” Z.
It is used with the two "dilution titration technique"

Calibration

techniques “DT Record calibration curve” and “DT Suppressors with
calibration curve” (with the modes CVS and CPVS). It should nor-
mally be around 0.5 and can be defined on the

Substances

tab of

the

EDIT WORKING METHOD PARAMETERS

window.

For the regression, all measurement points between 1 and the
measurement point after the evaluation ratio are used.

Note: The following condition must be fulfilled: „Begin of evalua-
tion“ > “Evaluation ratio” > „Addition ratio“.

Begin of evaluation

Only the measuring points between the „Begin of evaluation“ and
the „Evaluation ratio“ are used for the calculation of the regression
(linear, nonlinear and quadratic). It is used with the

Calibration

techniques DT Record calibration curve, DT Suppressors with cali-
bration curve , RC Sample with response curve and RC Record re-
sponse curve (with the modi CVS and CPVS). It should normally be
around 1.0 and can be defined on the

Substances

tab of the

EDIT

WORKING METHOD PARAMETERS

window.

Note: The following condition must be fulfilled: „Begin of evalua-
tion“ > “Evaluation ratio” > „Addition ratio“.