beautypg.com

Metrohm 797 VA Computrace User Manual

Page 227

background image

8.5 Determinations with voltammetric trace analysis

797 VA Computrace – Software

215

15. Follow the instructions in the appearing message windows.

Note: If you want to measure all samples with the same working
method, you can select

Repeat current method

for Working

method source on the

Automation

tab of the

GENERAL SETTINGS

window. The

SAMPLE TABLE

window is not accessible with

Repeat

current method

.

Perform VA determinations using the 838 Advanced Sample Processor

1.

Install the 838 Advanced Sample Processor (see Installation of
838 Advanced Sample Processor section 1.3
) and set a suitable
method (depending on the determination, see section 8.5
"Perform VA determinations using the 838 Advanced Sample
Processor"
, section 8.6 "Brightener Analysis with 838 Ad-
vanced Sample Processor and "MLAT"
, section 8.6 "Brightener
Analysis with 838 Advanced Sample Processor and "LAT"
, sec-
tion 8.6
"Suppressor Analysis with 838 Advanced Sample Proc-
essor and DT", section 8.6 "Suppressor Analysis with 838 Ad-
vanced Sample Processor and RC"
).

2.

Click on

MAIN WINDOW

/ Settings / General settings

and select

the

Hardware

tab.

3.

Define the

Sample processor

field of the

Hardware

tab.

4.

Modify the parameters for Automation in the

Automation

tab

of the

GENERAL SETTINGS

window as desired.

5.

If desired, test the automation parameters: Click on

MAIN WIN-

DOW

/ Settings / General settings

and select the

Automation

tab. Fill two sample vessels with water and place them on the

sample rack. Click on

, check the automation para-

meters and modify them if needed.

6.

Click on

or

MAIN WINDOW

/ Mode / Determination

.

7.

Click on

or

MAIN WINDOW

/

Window / Working method

specification

to open the

WORKING METHOD SPECIFICATIONS

window.

8.

Load the desired method into the

WORKING METHOD SPECIFI-

CATIONS

window (see How to Load a method, section 8.4).

9.

If desired, modify and save the loaded method (see How to
Edit the working method
, section 8.4).

10.

Click on

or

MAIN WINDOW / Window / Sample table

to

open the

SAMPLE TABLE

window.

11.

Load the desired sample table or edit the current sample table
(see Sample table, section 5.6).

12.

Transfer the desired sample amount into the sample vessels.
Place the sample vessels sequential on the sample rack of the