Glp diagnostics, Glp dummy cell tests – Metrohm 797 VA Computrace User Manual
Page 43
2.7 Settings menu
797 VA Computrace – Software
31
Note: Occasional electrode tests should be done
with
in the
COMPUTRACE CONTROL
window and not with the GLP Wizard (not to
overwrite GLP data).
Validation of a chosen method
Performs a validation of accuracy and precision us-
ing Standard Operating Procedures (SOP).
Note: When the GLP Wizard is executed, the current GLP data that
is stored in the software and in the 797 VA Computrace instrument
is overwritten.
GLP Diagnostics
1.
If you check (only) the
Diagnostics
checkbox on the
GLP Wizard
page of the
GLP WIZARD
window and press <
continue
>, you
get to the
Diagnostics
page of the
GLP WIZARD
window.
2.
Press <
continue
> one more time, and the
DIAGNOSTICS
win-
dow opens.
3.
In the
DIAGNOSTICS
window you can choose manually or with
the menu
select
which tests will be done. Press <
Start
> to get
the testing started.
4.
After completion you can save the data with
Save Report as
in
the
File
menu. You can print it with Print Report in the File
menu.
5.
Leave the
DIAGNOSTICS
window with
Exit
in the
File
menu.
6.
The
Summary of GLP validation
page of the
GLP WIZARD
win-
dow opens.
7.
Press <
Finish
>.
Note: Metrohm recommends to check all 4 procedures on the
GLP
Wizard
page of the
GLP WIZARD
window . The tests are progressed
in the order as listed. It is also possible to do the whole Diagnostics
procedure outside of the GLP Wizard: Click
Diagnostics
in
MAIN
WINDOW / HELP
to open the
DIAGNOSTICS
window.
GLP Dummy Cell Tests
1.
If you check (only) the
Dummy cell tests
checkbox on the
GLP
Wizard
page of the
GLP WIZARD
window and press <
continue
>,
you get to the
Dummy Cell Test_L
page of the
GLP WIZARD
window.
2.
Proceed as described and press <
continue
> one more time; the
MONITOR
window opens.