Validation of a chosen method

Performs a validation of accuracy and precision us-
ing Standard Operating Procedures (SOP).

Note: When the GLP Wizard is executed, the current GLP data that
is stored in the software and in the 797 VA Computrace instrument
is overwritten.

GLP Diagnostics

1.

If you check (only) the

Diagnostics

checkbox on the

GLP Wizard

page of the

GLP WIZARD

window and press <

continue

>, you

get to the

Diagnostics

page of the

GLP WIZARD

window.

2.

Press <

continue

> one more time, and the

DIAGNOSTICS

win-

dow opens.

3.

In the

DIAGNOSTICS

window you can choose manually or with

the menu

select

which tests will be done. Press <

Start

> to get

the testing started.

4.

After completion you can save the data with

Save Report as

in

the

File

menu. You can print it with Print Report in the File

menu.

5.

Leave the

DIAGNOSTICS

window with

Exit

in the

File

menu.

6.

The

Summary of GLP validation

page of the

GLP WIZARD

win-

dow opens.

7.

Press <

Finish

>.

Note: Metrohm recommends to check all 4 procedures on the

GLP

Wizard

page of the

GLP WIZARD

window . The tests are progressed

in the order as listed. It is also possible to do the whole Diagnostics
procedure outside of the GLP Wizard: Click

Diagnostics

in

MAIN

WINDOW / HELP

to open the

DIAGNOSTICS

window.

GLP Dummy Cell Tests

1.

If you check (only) the

Dummy cell tests

checkbox on the

GLP

Wizard

page of the

GLP WIZARD

window and press <

continue

>,

you get to the

Dummy Cell Test_L

page of the

GLP WIZARD

window.

2.

Proceed as described and press <

continue

> one more time; the

MONITOR

window opens.