Baseline – Metrohm 797 VA Computrace User Manual

Page 110

background image

5 Determination mode

797 VA Computrace – Software

98

Note: If you work with the modes CVS or CPVS the parameter list
is different.

Baseline

The

BASELINE

window contains the settings for baseline evaluation

for a single "variation" measurement (= all replications of a sample,
standard addition or calibration measurement) of a substance and
is opened by clicking the button for the selected substance in
the

Bsln.

column in the

Substances

tab of the

EDIT WORKING

METHOD PARAMETERS

window.

Auto. [ on, off ; on ]

Enable/disable the automatic peak evaluation.

Begin [ -5...+5 V ; 0 V ]

Manual setting of the start base point for baseline
evaluation. If

Auto

is enabled, the automatically

calculated start base point is displayed and the
field can not be edited.

End [ -5...+5 V ; 0 V ]

Manual setting of the end base point for baseline
evaluation. If

Auto

is enabled, the automatically

calculated end base point is displayed and the
field can not be edited.

Type [ Linear, Polynomial, Exponential , Horizontal ; Linear

]

Selection of baseline type. If

Auto

is enabled,

Line-

ar

is displayed and the field can not be edited.

Scope [ Whole peak, Front end, Rear end ; Whole peak ]

Selection of the range for baseline evaluation.

Can

only be edited with the

Type

“Linear”.

Switch to the previous page of this window.

Switch to the next page of this window.