6 Electroplating Bath VA
797 VA Computrace – Software
170
Note: If both
No. of conditioning measurements
[
1...100 ; 1 ] and
Auto Std.dev
(%) [ >0.1 ; 1 ] are
activated, the conditioning is finished as soon as
the first of the two criterias is reached.
Addition mixing time [ 0...80600 s ; 10 s ]
Time of mixing after Additions (for the first mea-
surement, the
Initial mixing time
is used).
Calibration curve [ path + file name ; ]
Define the calibration curve file (the file in which
the determination of the calibration was(will be)
saved).
The calibration curve can be recorded with
the
Calibration
technique “DT Record calibration
curve”.
Note: To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, the name of the calibration file defined
for
Calibration curve
, must match with the para-
meter
Sample identifier
on the
Determination
tab
with
Calibration
technique "DT Record calibration
curve".
Example for
Calibration curve
:
C:\User XYZ\Data\Calibration Suppressor.dth
Note: To make sure that for Suppressor analysis
with 838 Advanced Sample Processor and DT al-
ways the latest calibration file is taken for the cal-
culation, the path of the calibration file defined for
Calibration curve
, must match with the path de-
fined for the parameter
Data folder
(for the cur-
rently logged-in user) on the tab
User Directories
of the window
USER RIGHTS
.
Note: A determined calibration curve is only
adopted automatically if the checkbox
Save cali-
bration curves additionally without date and time
on the
General
tab of the
GENERAL SETTINGS
window is checked.
No. of replications [ 1...10 ; 2 ]
Number of replications (= total number of measu-
rements) for each variation (sample, standard ad-
dition, calibration level, conditioning measure-
ment). For cyclic modes (CV, CVS, CPVS) the “total
number of measurements” is the
No. of repli-
cations
multiplied with the number of
Save last ..