Result details with cvs and cpvs – Metrohm 797 VA Computrace User Manual
Page 203

6.3 Different settings and options with CVS and CPVS
797 VA Computrace – Software
191
Additional conditioning after sample transfer
(only active with
MLAT and RC Sample with response curve , and if
Initial electrode conditioning is activated)
If this option is activated, an additional electrode
conditioning is started after sample addition. This
option can only be activated, if Initial electrode
conditioning is activated.
Result details with CVS and CPVS
There is some additional information displayed in the
RESULTS
window with CVS and CPVS (compare to
Results
with other modes
section 5.5
):
Peak evaluation
(see also Peak evaluation with other modes, sec-
tion 5.5)
Q.Mean
Mean value of the evaluation charge for all repli-
cations of a variation.
Q.delta
Difference of two successive mean values of the
evaluation charge.
V (mL)
Volume added.
Q/Q(0)
Normalized evaluation charge.
Calibration data
(see also Calibration data with other modes, sec-
tion 5.5)
Calibr.
Calibration
techniques dt.rec.cc., dt.cc, mlat, lat,
rec.rc, smp.rc
(see Calibration techniques with
CVS and CPVS, section 6.2)
Sample data
(see also Sample data with other modes, section 5.5)
Intercept value
Measured value of the "Intercept solution" (see
Calibration techniques with CVS and CPVS, sec-
tion 6.2)
Substance evaluation
(see also Substance evaluation with other
modes, section 5.5)
Volume
With
Calibration
technique “DT Record calibration
curve”: Volume of added Suppressor standard so-
lution at the “Evaluation ratio”.
With
Calibration
technique “DT Suppressors with
calibration curve”: Volume of added Production
bath solution at the “Evaluation ratio”.