6.3 Different settings and options with CVS and CPVS

797 VA Computrace – Software

191

Additional conditioning after sample transfer

(only active with

MLAT and RC Sample with response curve , and if
Initial electrode conditioning is activated)
If this option is activated, an additional electrode
conditioning is started after sample addition. This
option can only be activated, if Initial electrode
conditioning is activated.

Result details with CVS and CPVS

There is some additional information displayed in the

RESULTS

window with CVS and CPVS (compare to

Results

with other modes

section 5.5

):

Peak evaluation

(see also Peak evaluation with other modes, sec-

tion 5.5)

Q.Mean

Mean value of the evaluation charge for all repli-
cations of a variation.

Q.delta

Difference of two successive mean values of the
evaluation charge.

V (mL)

Volume added.

Q/Q(0)

Normalized evaluation charge.

Calibration data

(see also Calibration data with other modes, sec-

tion 5.5)

Calibr.

Calibration

techniques dt.rec.cc., dt.cc, mlat, lat,

rec.rc, smp.rc

(see Calibration techniques with

CVS and CPVS, section 6.2)

Sample data

(see also Sample data with other modes, section 5.5)

Intercept value

Measured value of the "Intercept solution" (see
Calibration techniques with CVS and CPVS, sec-
tion 6.2
)

Substance evaluation

(see also Substance evaluation with other

modes, section 5.5)

Volume

With

Calibration

technique “DT Record calibration

curve”: Volume of added Suppressor standard so-
lution at the “Evaluation ratio”.
With

Calibration

technique “DT Suppressors with

calibration curve”: Volume of added Production
bath solution at the “Evaluation ratio”.