A U T O / S T E P M O D E S

4 - 1 1

*Note:

When using the AAMI or Custom test standard, only the applied

parts can be assigned; the Type cannot be changed.

• Make lead type assignments at the

Define Lead and Select Type

menu:

baw092f.eps

This selection allows the operator to define the Types for each of the
applied parts during the following tests: Mains on Applied Part,
Insulation Resistance – Applied Part, Patient Leakage, and Patient
Auxiliary Current. The Limits applied to the leads will be set based on
the Type selected.

Use the following procedure to test a defibrillator with two paddles, Type
BF, and three ECG leads, RA, LA and LL, all of which are Type CF.

Assuming that the ECG leads will be connected to the 601PRO applied
part terminals with the identical names, the paddles would have to be
connected to the RL and any V terminal. Define the lead types using the
following steps:

1. The first lead that appears for definition is RA. Since RA is Type CF,

press

TYPE CF

:

Define AP 1 and Select Type: RA BF

baw093f.eps

Press

enter

to advance to the next lead (RL).

2. Since RL is Type BF, press

TYPE BF

:

baw094f.eps


Press

ENT

to advance to the next lead (LA).