A U T O / S T E P M O D E S
4 - 1 1
*Note:
When using the AAMI or Custom test standard, only the applied
parts can be assigned; the Type cannot be changed.
• Make lead type assignments at the
Define Lead and Select Type
menu:
baw092f.eps
This selection allows the operator to define the Types for each of the
applied parts during the following tests: Mains on Applied Part,
Insulation Resistance – Applied Part, Patient Leakage, and Patient
Auxiliary Current. The Limits applied to the leads will be set based on
the Type selected.
Use the following procedure to test a defibrillator with two paddles, Type
BF, and three ECG leads, RA, LA and LL, all of which are Type CF.
Assuming that the ECG leads will be connected to the 601PRO applied
part terminals with the identical names, the paddles would have to be
connected to the RL and any V terminal. Define the lead types using the
following steps:
1. The first lead that appears for definition is RA. Since RA is Type CF,
press
TYPE CF
:
Define AP 1 and Select Type: RA BF
baw093f.eps
Press
enter
to advance to the next lead (RL).
2. Since RL is Type BF, press
TYPE BF
:
baw094f.eps
Press
ENT
to advance to the next lead (LA).