7. Configuring the Enclosure Leakage for the Auto Mode

Sequence

To take multiple

Enclosure Leakage

readings when performing tests in

Auto

mode:

• From the 601PRO

MAIN MENU

, press

SYSTEM SETUP

to access the

Select Setup Function

menu. Press

*MORE

until the following menu is

displayed, then press

ENCLOSURE LEAKAGE

:

baw015f.eps

• At the following menu, press

YES

.

baw016f.eps

Default Setting:

NO

• Press

enter

to save the setting and return to the

Select Setup Function

menu. Press

previous

to return to the

MAIN MENU

.

baw017f.eps

Thereafter, whenever an

Enclosure Leakage

test is performed as part of an

Auto

sequence, readings will be continually taken until

NEXT->

is pressed.

The last reading taken will then be used to determine Pass/Fail.

Note:

If running a custom standard auto sequence while multiple leakage

readings are enabled, the operator may experience a delay of up to

30 seconds when taking DC readings before the next test begins.

Refer to Chapter 4, Auto/Step Modes, for more information.