Fluke Biomedical 601PRO XL User Manual
Page 32

6 0 1 P R O S E R I E S
X L
2 - 8
7. Configuring the Enclosure Leakage for the Auto Mode
Sequence
To take multiple
Enclosure Leakage
readings when performing tests in
Auto
mode:
• From the 601PRO
MAIN MENU
, press
SYSTEM SETUP
to access the
Select Setup Function
menu. Press
*MORE
until the following menu is
displayed, then press
ENCLOSURE LEAKAGE
:
baw015f.eps
• At the following menu, press
YES
.
baw016f.eps
Default Setting:
NO
• Press
enter
to save the setting and return to the
Select Setup Function
menu. Press
previous
to return to the
MAIN MENU
.
baw017f.eps
Thereafter, whenever an
Enclosure Leakage
test is performed as part of an
Auto
sequence, readings will be continually taken until
NEXT->
is pressed.
The last reading taken will then be used to determine Pass/Fail.
☛
Note:
If running a custom standard auto sequence while multiple leakage
readings are enabled, the operator may experience a delay of up to
30 seconds when taking DC readings before the next test begins.
Refer to Chapter 4, Auto/Step Modes, for more information.