Fluke Biomedical 601PRO XL User Manual
Page 51

M A N U A L M O D E
3 - 1 1
4. Make lead type assignments at the
Define Lead and Select Type
menu:
baw039f.eps
This selection allows the operator to define the Types for each of the
applied parts during the following tests: Mains on Applied Part,
Insulation Resistance – Applied Part, Patient Leakage, and Patient
Auxiliary Current. The Limits applied to the leads will be set based on
the Type selected.
Use the following procedure to test a defibrillator with two paddles, Type
BF, and three ECG leads, RA, LA and LL, all of which are Type CF.
(This is an example of Condition 3 described on page 3-10.)
Assuming that the ECG leads will be connected to the 601PRO applied
part terminals with the identical names, the paddles would have to be
connected to the RL and any V terminal. Define the lead types using the
following steps:
1. The first lead that appears for definition is RA. Since RA is Type CF,
press
TYPE CF
:
baw040f.eps
Press
enter
to advance to the next lead (RL).
2. Since RL is Type BF, press
TYPE BF
:
baw041f.eps
Press
enter
to advance to the next lead (LA).