D E V I C E R E C O R D S

6 - 5

4. Device Information Record: Definition of Fields

Field

Name

Maximum

Characters

CONTROL#:

(15)**

PROCEDURE

ID:

(20)**

LOCATION: (30)**
DEVICE

TYPE: (20)**

MANUFACTURER: (25)**
SERIAL

#:

(16)**

TECHNICIAN:

Empty

field

(was

5)*

DATE OF LAST TEST:

Empty field (was 8)*

CLASS:

(2)**

TYPE:

(2)**

TEST STANDARD:

(10)**

601-PRO SERIAL#:

Empty field (was 15)*

LEADS:

(1)**

LEAD1

AP: (5)**

LEAD1

TYPE:

(2)**

LEAD2

AP: (5)**

LEAD2

TYPE:

(2)**

LEAD3

AP: (5)**

LEAD3

TYPE:

(2)**

LEAD4

AP: (5)**

LEAD4

TYPE:

(2)**

LEAD5

AP: (5)**

LEAD5

TYPE:

(2)**

POWER OFF DELAY:

(3)**

STARTUP

DELAY: (3)**

APPLIED

CURRENT:

(2)**

*These fields are present for compatibility with older versions of the

601PRO; and when received by the 601PRO Series

XL

, they are ignored. The

601PRO transmits them as empty fields.

**Restrictions, interpretation, and defaults for these fields are described
below:

⇒ Incoming device information records are checked for errors as

follows: The order, number, and size of each field and record are

fixed.

⇒ The Control Number must contain at least one character. The Test

Standard must contain a valid entry. All other fields may be empty.

⇒ Valid Classes are I, (space)I, II, or IP.

⇒ Valid Types and Lead Types are B, BF, CF, T and F.