Shortcut key 4: earth leakage test – Fluke Biomedical 601PRO XL User Manual
Page 62

6 0 1 P R O S E R I E S
X L
3 - 2 2
To perform a
Point-to-Point Resistance
test:
• Attach the RED lead to the RED input terminal and the BLACK lead to the
GREEN terminal.
• Place unknown R between the two leads.
• Press
START TEST (SOFT KEY 1)
.
Shortcut Key 4: Earth Leakage Test
Run an
Earth Leakage
test on the device being tested before performing any other
leakage tests.
Leakage current is measured the following ways:
♦ Earth Leakage Current, leakage current measured through DUT outlet Earth
♦ Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth),
leakage current measured through DUT outlet Earth
There is no need to attach a test lead; the 601PRO automatically connects the
measuring device internally.
The following outlet conditions apply when performing this test:
⇒ Normal Polarity, Earth Open, Outlet ON
⇒ Normal Polarity, Earth Open, L2 Open, Outlet ON
⇒ Reversed Polarity, Earth Open, Outlet ON
⇒ Reversed Polarity, Earth Open, L2 Open, Outlet ON
To perform the
Earth Leakage
test:
• From the
MAIN MENU
, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Attach the device's applied parts to the 601PRO applied part terminals if
applicable.
• Press
shortcut key 4
.The
Earth Leakage
test appears on the display, and the test
begins immediately:
baw055f.eps