beautypg.com

Shortcut key 4: earth leakage test – Fluke Biomedical 601PRO XL User Manual

Page 62

background image

6 0 1 P R O S E R I E S

X L

3 - 2 2

To perform a

Point-to-Point Resistance

test:

• Attach the RED lead to the RED input terminal and the BLACK lead to the

GREEN terminal.

• Place unknown R between the two leads.
• Press

START TEST (SOFT KEY 1)

.

Shortcut Key 4: Earth Leakage Test

Run an

Earth Leakage

test on the device being tested before performing any other

leakage tests.

Leakage current is measured the following ways:

♦ Earth Leakage Current, leakage current measured through DUT outlet Earth

♦ Earth Leakage Current AP-EARTH (ALL Applied Parts connected to Earth),

leakage current measured through DUT outlet Earth

There is no need to attach a test lead; the 601PRO automatically connects the

measuring device internally.

The following outlet conditions apply when performing this test:

⇒ Normal Polarity, Earth Open, Outlet ON

⇒ Normal Polarity, Earth Open, L2 Open, Outlet ON

⇒ Reversed Polarity, Earth Open, Outlet ON

⇒ Reversed Polarity, Earth Open, L2 Open, Outlet ON

To perform the

Earth Leakage

test:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Attach the device's applied parts to the 601PRO applied part terminals if

applicable.

• Press

shortcut key 4

.The

Earth Leakage

test appears on the display, and the test

begins immediately:

baw055f.eps