M A N U A L M O D E
3 - 2 9
• Press
SOFT KEYS 1-4
to select leakage tests:
⇒
SOFT KEY 1
toggles the DUT outlet Polarity from Normal to DUT Off to
Reverse Polarity, or Reverse to DUT Off to Normal Polarity.
⇒
SOFT KEY 2
toggles the DUT outlet from Earth to No Earth.
⇒
SOFT KEY 3
toggles the DUT outlet from L2 to No L2.
⇒
SOFT KEY 4
toggles the applied part from:
• RA to ALL (RMS)
• RL to ALL (RMS)
• LA to ALL (RMS)
• LL to ALL (RMS)
• V1-V6 to ALL (RMS)
• RA to ALL DC
• RL to ALL DC
• LA to ALL DC
• LL to ALL DC
• V1-V6 to ALL DC
☛
Note:
If the current test standard being used does not include Patient Auxiliary
Current DC readings, or the DC option is not enabled, then DC readings
will not be available through the
APPLIED PART SOFT KEY
selections.
Refer to Chapter 8, Standards and Principles.
• Press
APPLIED PART (SOFT KEY 4)
at any time to select the desired applied
part leakage current:
• Modify the configuration of the front panel outlet by pressing the appropriate
SOFT KEY
on the 601PRO:
• Press the
print data
key at any time to generate a printout of the latest
measurement.