M A N U A L M O D E

3 - 2 9

• Press

SOFT KEYS 1-4

to select leakage tests:

SOFT KEY 1

toggles the DUT outlet Polarity from Normal to DUT Off to

Reverse Polarity, or Reverse to DUT Off to Normal Polarity.

SOFT KEY 2

toggles the DUT outlet from Earth to No Earth.

SOFT KEY 3

toggles the DUT outlet from L2 to No L2.

SOFT KEY 4

toggles the applied part from:

• RA to ALL (RMS)

• RL to ALL (RMS)

• LA to ALL (RMS)

• LL to ALL (RMS)

• V1-V6 to ALL (RMS)

• RA to ALL DC

• RL to ALL DC

• LA to ALL DC

• LL to ALL DC

• V1-V6 to ALL DC


Note:

If the current test standard being used does not include Patient Auxiliary

Current DC readings, or the DC option is not enabled, then DC readings

will not be available through the

APPLIED PART SOFT KEY

selections.

Refer to Chapter 8, Standards and Principles.

• Press

APPLIED PART (SOFT KEY 4)

at any time to select the desired applied

part leakage current:

• Modify the configuration of the front panel outlet by pressing the appropriate

SOFT KEY

on the 601PRO:

• Press the

print data

key at any time to generate a printout of the latest

measurement.