Custom standard controlled power sequence, Modes – Fluke Biomedical 601PRO XL User Manual
Page 145

C U S T O M S T A N D A R D S
9 - 5
3. Custom Standard Controlled Power Sequence
The following table shows the controlled power sequence for custom test standards.
Startup and Power-off delays are performed, regardless of the tests enabled. Only the
tests enabled in the Custom Standard are performed. This sequence would be
performed in both
Auto
and
Step
modes.
IEC-601-1 STANDARD
DUT OUTLET
Description Polarity
Circuit
Line Voltage V
N/A
OFF
Protective Earth Resistance
Ω
N/A OFF
Insulation Resistance L1-L2-Case
Ω
N/A OFF
Insulation Resistance All-Case M
Ω
N/A OFF
Startup Delay (auto only)
Norm
Norm
Current in L2 - A
Norm
Norm
Accessible Voltage V
Norm
Norm
Enclosure Leakage - µA
Norm
Norm
Accessible Leakage - µA DC
Norm
Norm
Patient Leakage Current - µA
Norm
Norm
Patient Leakage Current - µA - DC Only
Norm
Norm
Mains on AP – µA
Norm
Norm
Patient Auxiliary - µA
Norm
Norm
Patient Auxiliary - µA - DC Only
Norm
Norm
Accessible Voltage V
Norm
No E
Earth Leakage - µA
Norm
No E
Enclosure Leakage - µA
Norm
No E
Accessible Leakage - µA DC
Norm
No E
Patient Leakage Current - µA
Norm
No E
Patient Leakage Current - µA - DC Only
Norm
No E
Patient Auxiliary - µA
Norm
No E
Patient Auxiliary - µA - DC Only
Norm
No E
Power OFF Delay
Norm
OFF
Startup Delay
Norm
No L2
Accessible Voltage V
Norm
No L2
Earth Leakage - µA
Norm
No L2
Enclosure Leakage - µA
Norm
No L2
Accessible Leakage - µA DC
Norm
No L2
Patient Leakage Current - µA
Norm
No L2
Patient Leakage Current - µA - DC Only
Norm
No L2
Patient Auxiliary - µA
Norm
No L2
Patient Auxiliary - µA - DC Only
Norm
No L2