C U S T O M S T A N D A R D S

9 - 5

3. Custom Standard Controlled Power Sequence

The following table shows the controlled power sequence for custom test standards.
Startup and Power-off delays are performed, regardless of the tests enabled. Only the
tests enabled in the Custom Standard are performed. This sequence would be
performed in both

Auto

and

Step

modes.

IEC-601-1 STANDARD

DUT OUTLET

Description Polarity

Circuit

Line Voltage V

N/A

OFF

Protective Earth Resistance

N/A OFF

Insulation Resistance L1-L2-Case

N/A OFF

Insulation Resistance All-Case M

N/A OFF

Startup Delay (auto only)

Norm

Norm

Current in L2 - A

Norm

Norm

Accessible Voltage V

Norm

Norm

Enclosure Leakage - µA

Norm

Norm

Accessible Leakage - µA DC

Norm

Norm

Patient Leakage Current - µA

Norm

Norm

Patient Leakage Current - µA - DC Only

Norm

Norm

Mains on AP – µA

Norm

Norm

Patient Auxiliary - µA

Norm

Norm

Patient Auxiliary - µA - DC Only

Norm

Norm

Accessible Voltage V

Norm

No E

Earth Leakage - µA

Norm

No E

Enclosure Leakage - µA

Norm

No E

Accessible Leakage - µA DC

Norm

No E

Patient Leakage Current - µA

Norm

No E

Patient Leakage Current - µA - DC Only

Norm

No E

Patient Auxiliary - µA

Norm

No E

Patient Auxiliary - µA - DC Only

Norm

No E

Power OFF Delay

Norm

OFF

Startup Delay

Norm

No L2

Accessible Voltage V

Norm

No L2

Earth Leakage - µA

Norm

No L2

Enclosure Leakage - µA

Norm

No L2

Accessible Leakage - µA DC

Norm

No L2

Patient Leakage Current - µA

Norm

No L2

Patient Leakage Current - µA - DC Only

Norm

No L2

Patient Auxiliary - µA

Norm

No L2

Patient Auxiliary - µA - DC Only

Norm

No L2