For the controlled power sequences, refer to Chapter 8, Standards and
Principles, pages 8-5 and 8-6 for IEC 601 test limits, and page 8-12 for HEI 95
test limits.
Although the conventional test sequences are performed in a different order,
the results and printed output are in the same order as the controlled power
sequences. See Chapter 8, pages 8-7 through 8-9 for IEC 601 test limits, and
page 8-13 for HEI 95 test limits.
The 601PRO, not the device record, determines which sequence is used. The
601PRO SERIES
XL
units are supplied with the controlled power sequences
disabled.
☛
Note:
The setup options that are defined are
not
saved with the
device records.
Startup and Power Off Delays have been inserted into the conventional test
sequences to allow customization of the sequence. Long or short delays can be
programmed into a device record, considering the specific device under test.
The Power Off Delay, for all sequences, has a range of 1 to 9999 seconds. The
Startup Delay has a range of 0 to 9999 seconds.
☛
Note: In
Auto/Step
mode, Startup and Power Off delays are
performed, regardless of the tests enabled.