N O T I C E S , W A R N I N G S A N D C O N T E N T

v i i

Hazard Warnings

Warning! Power Rating. The 601PRO's mains power input must be connected to a

power receptacle that provides voltage within the specified rating for the system.
Connection must be made via the Mains Power cord provided by Fluke Biomedical.

Use of an incompatible power receptacle or incorrect Mains Power cord may produce
electrical shock and fire hazards. Acceptable Mains Voltage ranges are 90~VAC to
132~VAC, and 180~VAC to 240~VAC 50/60 Hz. The current ratings for the 601PRO are
as follows:

Europe:

<=15A (Fused by 15A circuit breaker)

United Kingdom:

<=13A (Fused by 13A fused Mains Power Cord)

Australia:

<=10A (Fused by 10A circuit breaker)

Warning! Internal Voltage. Always turn off the power switch and unplug the power

cord before cleaning the 601PRO's outer surface.

Warning! Liquids. Avoid spilling liquids on the analyzer; fluid seepage into internal

components creates a potential shock hazard. Do not operate the instrument if internal
components are exposed to fluid.

Precautions


The following precautions are provided to help you avoid damaging the system:

To power up the 601PRO, place the index finger on the rocker switch and use a

rolling motion to push from "OFF" to "ON." Do NOT forcefully push or snap the
rocker switch. This may cause the unit to shut off.


Caution: Service. The 601PRO should be serviced by authorized service personnel.

Only qualified technical personnel should perform troubleshooting and service
procedures on internal components.


Caution: Environmental Conditions. Do not expose the system to temperature

extremes. Ambient temperatures should remain between 18-40°C. System
performance may be adversely affected if temperatures fluctuate above or below
this range.

Caution: Do Not Immerse. Clean only with a mild detergent, and wipe down

with a gentle cloth.