6 0 1 P R O S E R I E S

X L

3 - 3 0

Shortcut Key 9: IEC 1010 Accessible Voltage/Leakage Test

The Accessible Voltage and Leakage Tests are both accessible using Shortcut Key 9.

The tests cannot be selected from the

Manual

menu unless the IEC 1010 test standard

is selected.

Important:

The Accessible Voltage and Leakage test uses the IEC 1010 Test Load

exclusively. This is true no matter what the currently defined test

standard is.

Accessible Voltage

Accessible Voltage measurements are taken from the RED input lead referenced to

the Protective Earth terminal on the Mains Inlet. Accessible Voltage cannot be

selected from the

Manual

menu unless the IEC 1010 test standard is selected.

The following outlet conditions apply when performing this test:

⇒ Normal Polarity, Earth Open, Outlet ON

⇒ Normal Polarity, Outlet ON

⇒ Normal Polarity, L2 Open, Outlet ON

⇒ Reversed Polarity, Outlet ON

⇒ Reversed Polarity, Earth Open, Outlet ON

⇒ Reversed Polarity, L2 Open, Outlet ON

To measure Accessible Voltage using the

test-shortcut

keys:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Attach the RED lead to exposed metal on the enclosure of the DUT.
• Press

shortcut key 9

. The

Accessible Voltage

test appears on the display, and

the test begins immediately. Display values are continuously updated until

another test is selected.