S P E C I F I C A T I O N S
A - 5
Insulation Resistance
Configuration:
Measurement taken through the DUT outlet as well
as the Patient leads
Applied Voltage:
500 V ± 10% through a 5 M
Ω
limiting resistor
Accuracy:
± 5% of reading ± 2 LSD
Measurement Range:
0.5-400.0 M
Ω
Current Consumption
Configuration:
DUT outlet can be configured as Open or Closed
Earth and L2 and Reverse/Normal polarity
Range:
0.0-15.0 Amps AC true RMS
Accuracy:
± 5% of reading ± 2 LSD
ECG Simulation and Performance Testing
Accuracy:
± 2% of reading for rate or ± 5% of reading for
Amplitude, which is fixed at 1 mV peak on a
LEAD II ECG connection (except for triangle wave,
which is 2 mV peak to peak)
Square wave:
0.125, 2 Hz 50% duty cycle
Sine wave:
10, 40, 50, 60, 100 Hz
Triangle wave:
2 Hz, 2 mV
ECG complex:
30, 60, 120, 180, 240 BPM
Pulse:
30,60 BPM 63 ms pulse width
600-700 µs rise and fall time
Other waveforms:
A-Fib, A-Flutter, A-Tach, Idioventricular, PVC1, R-
on-T, Run, V-Fib, V-Tach