Insulation resistance, Current consumption, Ecg simulation and performance testing – Fluke Biomedical 601PRO XL User Manual

Page 169

background image

S P E C I F I C A T I O N S

A - 5

Insulation Resistance

Configuration:

Measurement taken through the DUT outlet as well

as the Patient leads

Applied Voltage:

500 V ± 10% through a 5 M

limiting resistor

Accuracy:

± 5% of reading ± 2 LSD

Measurement Range:

0.5-400.0 M

Current Consumption

Configuration:

DUT outlet can be configured as Open or Closed

Earth and L2 and Reverse/Normal polarity

Range:

0.0-15.0 Amps AC true RMS

Accuracy:

± 5% of reading ± 2 LSD

ECG Simulation and Performance Testing

Accuracy:

± 2% of reading for rate or ± 5% of reading for

Amplitude, which is fixed at 1 mV peak on a

LEAD II ECG connection (except for triangle wave,

which is 2 mV peak to peak)

Square wave:

0.125, 2 Hz 50% duty cycle

Sine wave:

10, 40, 50, 60, 100 Hz

Triangle wave:

2 Hz, 2 mV

ECG complex:

30, 60, 120, 180, 240 BPM

Pulse:

30,60 BPM 63 ms pulse width

600-700 µs rise and fall time

Other waveforms:

A-Fib, A-Flutter, A-Tach, Idioventricular, PVC1, R-

on-T, Run, V-Fib, V-Tach