Fluke Biomedical 601PRO XL User Manual
Page 121

S T A N D A R D S A N D P R I N C I P L E S
8 - 7
IEC 601-1 Conventional Test Sequence and Test Limits for Auto and Step Tests
DC-Only tests will only be performed if DC readings are enabled. The types (B, BF, and CF) represent
applied parts for Insulation Resistance, Patient Leakage Current, Patient Auxiliary, and Mains on Applied
Part.
IEC 601-1 STANDARD
DUT OUTLET
CLASS I
CLASS II
CLASS IP
DESCRIPTION
POLARITY
CIRCUIT
B BF CF B BF CF B BF CF
Line Voltage V
N/A
OFF
NL
NL
NL NL NL NL NT NT NT
Protective Earth
Resistance
Ω
**
N/A
OFF 0.2 0.2 0.2 NT NT NT NT NT NT
Insulation Resistance L1-
L2- Case M
Ω
***
N/A
OFF 2 2 2 NT NT NT NT NT NT
Insulation Resistance All
Case M
Ω
N/A
OFF NT 2 2 NT NT NT NT NT NT
Startup Delay (Auto only)
Norm
Norm
N/A
N/A
N/A
Current in L2 -- A
Norm
Norm
NL
NL
NL
NL
NL
NL
NT
NT
NT
Earth Leakage --
µA
Norm
No
E 500 500 500 NT NT NT NT NT NT
Power Off Delay
Norm
No L2
No E
N/A
N/A
N/A
N/A
N/A
N/A
Earth Leakage --
µA
Norm No
L2
No E
1000
1000
1000 NT NT NT NT NT NT
Startup Delay (Auto only)
Reverse
No E
N/A
N/A
N/A
N/A
N/A
N/A
Earth Leakage --
µA
Reverse No
E 500 500 500 NT NT NT NT NT NT
Power Off Delay
Reverse
No L2
No E
N/A
N/A
N/A
N/A
N/A
N/A
Earth Leakage --
µA
Reverse No
L2
No E
1000
1000
1000 NT NT NT NT NT NT
Startup Delay (Auto only)
Norm
Norm
N/A
N/A
N/A
N/A
N/A
N/A
Enclosure Leakage --
µA
Norm
Norm 100 100 100 100 100 100 NT
NT NT
Power Off Delay
Norm
No L2
N/A
N/A
N/A
Enclosure Leakage --
µA
Norm
No
L2 500 500 500 500 500 500 NT NT NT
Startup Delay
Norm
No E
N/A
N/A
N/A
N/A
N/A
N/A
Enclosure Leakage --
µA
Norm
No
E 500 500 500 NT NT NT NT NT NT
Power Off Delay
N/A
OFF
N/A
N/A
N/A
Startup Delay (Auto only)
Reverse
Norm
N/A
N/A
N/A
Enclosure Leakage --
µA
Reverse Norm 100 100 100 100 100 100 NT NT NT
Power Off Delay
Reverse
No L2
N/A
N/A
N/A
Enclosure Leakage --
µA
Reverse No
L2 500 500 500 500 500 500 NT NT NT
Startup Delay
Reverse
No E
N/A
N/A
N/A
N/A
N/A
N/A
Enclosure Leakage --
µA
Reverse No
E 500 500 500 NT NT NT NT NT NT
Power Off Delay
N/A
OFF
N/A
N/A
N/A
Startup Delay (Auto only)
Norm
Norm
N/A
N/A
N/A
Patient Leakage Current -
µA
Norm
Norm 100 100 10 100 100 10 100 100 10
Power Off Delay
Norm
No L2
N/A
N/A
N/A
Patient Leakage Current -
µA
Norm No
L2 500 500 50 500 500 50 NT NT NT
Startup Delay
Norm
No E
N/A
N/A
N/A
N/A
N/A
N/A
Patient Leakage Current -
µA
Norm
No
E 500 500 50 NT NT NT NT NT NT
Power Off Delay
N/A
OFF
N/A
N/A
N/A
Startup Delay (Auto only)
Reverse
Norm
N/A
N/A
N/A
Patient Leakage Current -
µA
Reverse Norm 100 100 10 100 100 10 NT NT NT
Power Off Delay
Reverse
No L2
N/A
N/A
N/A
Patient Leakage Current -
µA
Reverse No
L2 500 500 50 500 500 50 NT NT NT