S T A N D A R D S A N D P R I N C I P L E S

8 - 7

IEC 601-1 Conventional Test Sequence and Test Limits for Auto and Step Tests

DC-Only tests will only be performed if DC readings are enabled. The types (B, BF, and CF) represent
applied parts for Insulation Resistance, Patient Leakage Current, Patient Auxiliary, and Mains on Applied
Part.

IEC 601-1 STANDARD

DUT OUTLET

CLASS I

CLASS II

CLASS IP

DESCRIPTION

POLARITY

CIRCUIT

B BF CF B BF CF B BF CF

Line Voltage V

N/A

OFF

NL

NL

NL NL NL NL NT NT NT

Protective Earth
Resistance

**

N/A

OFF 0.2 0.2 0.2 NT NT NT NT NT NT

Insulation Resistance L1-
L2- Case M

***

N/A

OFF 2 2 2 NT NT NT NT NT NT

Insulation Resistance All
Case M

N/A

OFF NT 2 2 NT NT NT NT NT NT

Startup Delay (Auto only)

Norm

Norm

N/A

N/A

N/A

Current in L2 -- A

Norm

Norm

NL

NL

NL

NL

NL

NL

NT

NT

NT

Earth Leakage --

µA

Norm

No

E 500 500 500 NT NT NT NT NT NT

Power Off Delay

Norm

No L2

No E

N/A

N/A

N/A

N/A

N/A

N/A

Earth Leakage --

µA

Norm No

L2

No E

1000

1000

1000 NT NT NT NT NT NT

Startup Delay (Auto only)

Reverse

No E

N/A

N/A

N/A

N/A

N/A

N/A

Earth Leakage --

µA

Reverse No

E 500 500 500 NT NT NT NT NT NT

Power Off Delay

Reverse

No L2

No E

N/A

N/A

N/A

N/A

N/A

N/A

Earth Leakage --

µA

Reverse No

L2

No E

1000

1000

1000 NT NT NT NT NT NT

Startup Delay (Auto only)

Norm

Norm

N/A

N/A

N/A

N/A

N/A

N/A

Enclosure Leakage --

µA

Norm

Norm 100 100 100 100 100 100 NT

NT NT

Power Off Delay

Norm

No L2

N/A

N/A

N/A

Enclosure Leakage --

µA

Norm

No

L2 500 500 500 500 500 500 NT NT NT

Startup Delay

Norm

No E

N/A

N/A

N/A

N/A

N/A

N/A

Enclosure Leakage --

µA

Norm

No

E 500 500 500 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Reverse

Norm

N/A

N/A

N/A

Enclosure Leakage --

µA

Reverse Norm 100 100 100 100 100 100 NT NT NT

Power Off Delay

Reverse

No L2

N/A

N/A

N/A

Enclosure Leakage --

µA

Reverse No

L2 500 500 500 500 500 500 NT NT NT

Startup Delay

Reverse

No E

N/A

N/A

N/A

N/A

N/A

N/A

Enclosure Leakage --

µA

Reverse No

E 500 500 500 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Norm

Norm

N/A

N/A

N/A

Patient Leakage Current -
µA

Norm

Norm 100 100 10 100 100 10 100 100 10

Power Off Delay

Norm

No L2

N/A

N/A

N/A

Patient Leakage Current -
µA

Norm No

L2 500 500 50 500 500 50 NT NT NT

Startup Delay

Norm

No E

N/A

N/A

N/A

N/A

N/A

N/A

Patient Leakage Current -
µA

Norm

No

E 500 500 50 NT NT NT NT NT NT

Power Off Delay

N/A

OFF

N/A

N/A

N/A

Startup Delay (Auto only)

Reverse

Norm

N/A

N/A

N/A

Patient Leakage Current -
µA

Reverse Norm 100 100 10 100 100 10 NT NT NT

Power Off Delay

Reverse

No L2

N/A

N/A

N/A

Patient Leakage Current -
µA

Reverse No

L2 500 500 50 500 500 50 NT NT NT