beautypg.com

Portable devices, Portable devices located in isolated power systems, Testing three-phase portable devices – Fluke Biomedical 601PRO XL User Manual

Page 112

background image

6 0 1 P R O S E R I E S

X L

7 - 2

9. Press the

Start Test

key to initiate the test. This is a point-to-point resistance

measurement.

10. Make sure the reading is below 0.1 ohms.

11. Remove the lead connected to the GREEN jack and set aside.

12. Press the

Enclosure

Leakage

key to run the

Enclosure Leakage Current

test.

13. Make sure the 601PRO front panel outlet is normal polarity, earthed, and L2 for

appropriate limits.

14. Repeat the

Enclosure Leakage Current

test with the device under test on.

2. Portable Devices

Test the portable device as detailed in Chapter 3, Manual Mode, using procedures for

Manual

,

Automatic

, or

Step

modes.

3. Portable Devices Located in Isolated Power Systems

Portable devices located in isolated power systems should be tested on an earth-

referenced power system.

Remove the DUT to an area with an earth-referenced power system. Or use an

extension cord to bring earth-referenced power to the DUT.

4. Testing Three-Phase Portable Devices

When testing three-phase portable devices, the operator must be able to interrupt

earth, L2, and reverse polarity. Three-phase equipment cannot be plugged directly

into the 601PRO; an adapter that interrupts the earth, L2 and reverses polarity must

be used.

1. Plug the device to be tested into an adapter that allows outlet configuration.

2. Plug the 601PRO into a nearby outlet or connect a cable from the protective earth

connector on the back of the 601Pro to conductive surface with the same earth

point as the device under test.

3. Connect a test lead from the RED input jack on the 601PRO to the enclosure

(case) of the device.

4. Follow the

Manual

mode test procedures in Chapter 3. The adapter should be

configured the same as the 601PRO for appropriate limits.