Dc only, Voltage measurements – main, single, and dual lead, Protective earth resistance – Fluke Biomedical 601PRO XL User Manual
Page 166

6 0 1 P R O S E R I E S
X L
A - 2
DC Only
IEC-601/AAMI
IEC-1010
Measurement Range:
0-8,000 µA
0-16,000 µA*
Accuracy:
± 1% of reading ± 1µA (@
DC)
±2 % of reading ± 2 µA (@
DC)
Input Impedance:
1000
Ω
± 1% (@ DC)
2000
Ω
± 1% (@ DC)
Cutoff Frequency:
5 Hz (approximate)
5 Hz (approximate)
50 Hz Rejection:
60 dB minimum
60 dB minimum
For the Equivalent Device, Equivalent Patient, and Mains on Applied Part Leakage
Current tests, the following exceptions apply:
Accuracy:
± 2% of reading ± 6 µA
Applied Voltage:
≥
110% of mains
*For some leakage tests, the displayed range may be 0.0-9999 when using the IEC
1010 test load.
Voltage Measurements – Main, Single, and Dual Lead
Configuration:
Single Lead: Red input jack to mains earth
Mains: L1 to L2, L1 to E, L2 to E
Dual Lead: Between red and black input jacks not
referenced to earth ground
Measurement Range:
0.000-300.0 V true RMS
Accuracy:
DC -100 Hz ± 1.5% of reading ± 1 LSD
Input Impedance:
≥
0.5 M
Ω
Protective Earth Resistance
The following conditions apply:
•
Selectable test current for 1 A, 10 A, or 25 A with open circuit voltage less than 6
VAC RMS
•
Open circuit resistance is defined as greater than 10
Ω
• Single lead measurement is to be taken between the RED terminal and the
receptacle FE terminal.