Dc only, Voltage measurements – main, single, and dual lead, Protective earth resistance – Fluke Biomedical 601PRO XL User Manual

Page 166

background image

6 0 1 P R O S E R I E S

X L

A - 2

DC Only

IEC-601/AAMI

IEC-1010

Measurement Range:

0-8,000 µA

0-16,000 µA*

Accuracy:

± 1% of reading ± 1µA (@
DC)

±2 % of reading ± 2 µA (@
DC)

Input Impedance:

1000

± 1% (@ DC)

2000

± 1% (@ DC)

Cutoff Frequency:

5 Hz (approximate)

5 Hz (approximate)

50 Hz Rejection:

60 dB minimum

60 dB minimum

For the Equivalent Device, Equivalent Patient, and Mains on Applied Part Leakage
Current tests, the following exceptions apply:

Accuracy:

± 2% of reading ± 6 µA

Applied Voltage:

110% of mains


*For some leakage tests, the displayed range may be 0.0-9999 when using the IEC
1010 test load.

Voltage Measurements – Main, Single, and Dual Lead

Configuration:

Single Lead: Red input jack to mains earth

Mains: L1 to L2, L1 to E, L2 to E

Dual Lead: Between red and black input jacks not
referenced to earth ground

Measurement Range:

0.000-300.0 V true RMS

Accuracy:

DC -100 Hz ± 1.5% of reading ± 1 LSD

Input Impedance:

0.5 M

Protective Earth Resistance

The following conditions apply:

Selectable test current for 1 A, 10 A, or 25 A with open circuit voltage less than 6
VAC RMS

Open circuit resistance is defined as greater than 10

• Single lead measurement is to be taken between the RED terminal and the

receptacle FE terminal.