Test principle: vde, equivalent device leakage – Fluke Biomedical 601PRO XL User Manual
Page 137

S T A N D A R D S A N D P R I N C I P L E S
8 - 2 3
Test Principle: VDE, Equivalent Device Leakage
Note:
S3 = Closed
baw145f.eps
Function:
Measures the current that flows between the mains terminals of the DUT to the PE, applied parts, and
exposed conductive parts of the DUT in response to an isolated AC voltage source with a magnitude of
at least 110% of mains voltage.
Applicable to:
VDE 751.1
VDE
701