Fluke Biomedical 601PRO XL User Manual
Page 126

6 0 1 P R O S E R I E S
X L
8 - 1 2
HEI 95 Controlled Power Sequence and Test Limits for Auto and Step Tests
DC-Only tests will only be performed if DC readings are enabled. The types (B, BF, and CF) represent
applied parts for Insulation Resistance, Patient Leakage Current, Patient Auxiliary, and Mains on Applied
Part.
HEI-95 STANDARD
DUT OUTLET
CLASS I
CLASS II
DESCRIPTION POLARITY CIRCUIT
B
BF CF
B
BF
CF
Protective Earth
Resistance
Ω
N/A OFF
0.2
0.2
0.2 NT NT NT
Insulation
Resistance L1-L2-
Case M
Ω
N/A
OFF 50 50 50 NT NT NT
Insulation
Resistance All-Case
M
Ω
N/A OFF
NT
50
50
NT
NT
NT
Startup Delay (Auto
only)
Norm
Norm
Enclosure Leakage
µA
Norm
Norm 500 500 500 100 100 10
Patient Leakage
Current µA
Norm Norm 500
500 50 100 100 10
Patient Leakage
Current µA – DC
Only
Norm Norm 500
500 50 100 100 10
Earth Leakage
µA
Norm
No
E 500 500 500 NT NT NT
Enclosure Leakage
µA
Norm
No
E 500 500 500 NT NT NT
Patient Leakage
Current µA
Norm No
E
500
500
50
NT
NT
NT
Patient Leakage
Current µA – DC
Only
Norm No
E
500
500
50
NT
NT
NT
Power Off Delay
Norm
No L2
Startup Delay (Auto
only)
Reverse
Norm
Enclosure Leakage
µA
Reverse Norm 500 500 500 100 100 10
Patient Leakage
Current µA
Reverse Norm 500
500 50 100 100 10
Patient Leakage
Current µA – DC
Only
Reverse Norm 500
500 50 100 100 10
Earth Leakage µA
Reverse
No E
500
500
500
NT
NT
NT
Enclosure Leakage
µA
Reverse No
E 500 500 500 NT NT NT
Patient Leakage
Current µA
Reverse No
E 500
500
50 NT NT NT
Patient Leakage
Current µA – DC
Only
Reverse No
E 500
500
50 NT NT NT