Custom standard controlled sequence (cont.) – Fluke Biomedical 601PRO XL User Manual

Page 147

background image

C U S T O M S T A N D A R D S

9 - 7

Custom Standard Controlled Sequence (Cont.)

IEC-601-1 STANDARD

DUT OUTLET

Description Polarity

Circuit

Line Voltage V

N/A

OFF

Protective Earth Resistance

N/A OFF

Insulation Resistance L1-L2-Case

N/A OFF

Insulation Resistance All-Case M

N/A OFF

Startup Delay (auto only)

Norm

Norm

Current in L2 - A

Norm

Norm

Accessible Voltage V

Norm

Norm

Enclosure Leakage - µA

Norm

Norm

Accessible Leakage - µA DC

Norm

Norm

Patient Leakage Current - µA

Norm

Norm

Patient Leakage Current - µA - DC Only

Norm

Norm

Mains on AP – µA

Norm

Norm

Patient Auxiliary - µA

Norm

Norm

Patient Auxiliary - µA - DC Only

Norm

Norm

Accessible Voltage V

Norm

No E

Earth Leakage - µA

Norm

No E

Enclosure Leakage - µA

Norm

No E

Accessible Leakage - µA DC

Norm

No E

Patient Leakage Current - µA

Norm

No E

Patient Leakage Current - µA - DC Only

Norm

No E

Patient Auxiliary - µA

Norm

No E

Patient Auxiliary - µA - DC Only

Norm

No E

Power OFF Delay

Norm

OFF

Startup Delay

Norm

No L2

Accessible Voltage V

Norm

No L2

Earth Leakage - µA

Norm

No L2

Enclosure Leakage - µA

Norm

No L2

Accessible Leakage - µA DC

Norm

No L2

Patient Leakage Current - µA

Norm

No L2

Patient Leakage Current - µA - DC Only

Norm

No L2

Patient Auxiliary - µA

Norm

No L2

Patient Auxiliary - µA - DC Only

Norm

No L2