2. The Power-Up Sequence

baw133f.eps

The 601PRO immediately begins a self-check routine at power-up while the

following instrument identification screen is displayed:

baw003f.eps

Power-up tests include internal self-checks and wall outlet checks. If any of the

power-up tests fail, the 601PRO displays an error message. Refer to Chapter 11,

Error Messages, Troubleshooting, and Support, for additional information.

When diagnostics are completed, the

MAIN MENU

will appear

:

baw004f.eps

To repeat the self-check routine at any time, press

UTILITIES

from the

MAIN MENU

,

then

SYSTEM TEST

.

Note:

At power-up, the 601PRO determines the polarity of the Mains power. If

the Mains polarity is normal (normal for a grounded neutral system), L2 in

the DUT Outlet will be opened whenever a NO L2 condition exists, but if

the Mains polarity is reversed, L1 in the DUT Outlet will be opened

whenever a NO L2 condition exists. This ensures that the highest Mains

voltage remains connected to the device under test for a NO L2 single-

fault condition. The operator will not be notified at power-up if the Mains

Voltage is reversed.

Note:

At power-up, the number of applied parts will be set to zero (0) for all

standards except VDE 751, which will default to five (5) applied parts.

The current source will default to 1 Amp.