M A N U A L M O D E

3 - 3 3

The limit value and the class/type designations will vary depending on the currently

selected test standard.

The following test conditions apply:

⇒ Mains supply to DUT L1 and L2 Open

⇒ DUT L1 and L2 shorted together

⇒ 110% of mains voltage is applied between L1, L2 and Earth at the DUT

outlet.

To perform an Equivalent Device Leakage test:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Press the

/

shortcut key

. The

Equivalent Device Leakage

test is displayed:

baw063f.eps

• Press

START TEST (SOFT KEY 1)

to begin the test.

WARNING:

High voltage is present at the DUT terminals while measurements are

being taken.

• Press the

print data

key to generate a printout of the latest measurement.

To perform a calibration from the

Equivalent Device Leakage

test, press

CALIBRATE (SOFT KEY 2)

.

• Disconnect ALL patient leads, test leads, and DUT outlet connections.

• Press

CAL

to begin calibration, as shown

.

baw059f.eps