T E S T I N G D E V I C E S

7 - 3

5. Testing Conductive Surfaces

1. Connect the RED test lead to the RED input terminal, and the BLACK test lead

to the BLACK input terminal.

2. Press the

TESTS/AUTOMODES

key on the 601PRO.

3. Press the

MANUAL

key, then press

*MORE

until the

DUAL LEAD LEAKAGE

menu is displayed.

4. Press

DUAL LEAD LEAKAGE

. The following menu is displayed:

baw111f.eps

5. Measure point-to-point leakage between the RED and BLACK input jacks on the

601PRO.

6. Detachable Power Supply Cable

The 601PRO has an isolated measuring circuit that can measure differences between

two points, independently of earth points.

1. Power up the 601PRO, and plug the detachable cable into the 601PRO Front

Panel.

2. Connect the test lead from the RED input jack to earth of the detachable power

supply cable.

3. Press the

Earth Resistance

(

shortcut key 3

), and then select the test current as

described in Chapter 3.

4. Press the

Earth Resistance

key once more to apply current.

The detachable power supply cable resistance should be less than 0.1 ohms.

7. Battery-Powered Equipment

1. Power up the 601PRO and connect a test lead from the GREEN input jack of the

601PRO to the enclosure (case) of the DUT.

2. Connect patient-applied leads to the applied part terminals on the 601PRO. Refer

to the

Automatic

mode procedures in Chapter 4, Auto/Step Modes.

3. Power up the device under test.

4. Select Class IP (internally powered) and run an auto-sequence.