Fluke Biomedical 601PRO XL User Manual
Page 167

S P E C I F I C A T I O N S
A - 3
The selectable test current for 1A, 10A
,
or 25A with open circuit voltage is less
than 6 VAC RMS, where open circuit resistance is defined as greater than 10
Ω
.
Single lead measurement is to be taken between the red terminal and the
receptacle FE terminal.
Range: 0.000-2.999
Ω
Accuracy:
± 5 % of reading ± 4 m
Ω
[see (a), (b), (c), and (d)
below]
Open Circuit Resistance:
≥
10
Ω
Current Source Output:
1 Amp selection
0.85-1.15 Amps AC RMS (0.000-2.999
Ω
)
10 Amp selection
9.50-10.50 Amps AC RMS (0.000-0.999
Ω
)
2.00-10.50 Amps AC RMS (1.000-2.999
Ω
)
25 Amp selection
22.5-25 Amps AC RMS (0.000-0.200
Ω
)
15.0-25 Amps AC RMS (0.000-0.300
Ω
)
2.0-25 Amps AC RMS (0.000-2.999
Ω
)
(a) Accuracy specification applies to measurements with 1 Amp and 25 Amp test
current where the load inductance is less than 60 µH and the phase angle is less
than 10 degrees. The Calibration resistance is required to be less than 35 m
Ω
(factory-supplied test leads in good condition).
Accuracy specification applies to measurements with 10 Amp test current where
the load inductance is less than 30 µH and the phase angle is less than 10
degrees. The calibration resistance is required to be less than 35 m
Ω
(factory-
supplied test leads in good condition).
(b) Specified accuracy with 10 Amp test current for load inductance up to 60 µH and
phase angle greater than 10 degrees:
5% of reading ± 20 m
Ω
(resistance range of 10-199 m
Ω
)
5% of reading ± 10 m
Ω
(resistance range of 200-2999 m
Ω
)