Setting up the 601pro, Using factory default settings – Fluke Biomedical 601PRO XL User Manual

Page 25

background image

2-1

Setting Up the 601PRO

This chapter describes the factory default settings and how to change them,

and provides instructions for one-time customization of the 601PROXL using

the

SYSTEM SETUP

key and the

Select Setup Function

menu.

1. Using Factory Default Settings
2. Selecting the Test Standard
3. Selecting the Printer Output
4. Selecting the RS232 Baud Rate
5. Activating the Beeper
6. Setting the Time and Date
7. Configuring the Enclosure Leakage for the Auto

Mode Sequence

8. Selecting Language Options
9. Selecting the DC Option
10. Selecting the Auto/Step Tests: Controlled Power

Sequences or 601CE Conventional Test

Sequences

11. Enabling Stop on Failure
12. Configuring for Device Records or Templates

1. Using Factory Default Settings

The factory default settings are shown below.

Description

Factory Default Setting

Test Standard

IEC 601-1 (Class I, BF)

Number of Applied Parts

0 (To change, refer to Chapter 3)

Printer Output

Internal

RS232 9600

Baud

Beeper On
Time Format

24 hour

Date Format

MDY

Enclosure Leakage (multiple readings
during the Auto Mode sequence)

Off

Keyboard Language

English

Display/Print Language

English

DC Readings

Off

Auto/Step Sequences: Controlled Power
Sequence

On

Stop on Failure

Off

DEV RECS / TEMPLATES

DEVICE REC

Chapter

2