Lead type definitions – Fluke Biomedical 601PRO XL User Manual
Page 49

M A N U A L M O D E
3 - 9
To choose the current source to be used for the Protective Earth Resistance test, press
the
view present settings
key, and press
AMPERES
at the following menu:
baw036f.eps
The test current displayed in the upper right corner will toggle between the available
currents.
• Press enter to return to the
MAIN MENU
.
Lead Type Definitions
The
CLASS/TYPE
selection sets the Type of EQUIPMENT that is being
tested. This affects the limits for the Protective Earth Resistance, Insulation
Resistance L1, L2-Case, Earth Leakage, and Enclosure Leakage Tests.
When Applied Part Types are defined, this affects the limits as well as which
tests are being performed on each of the individual leads. The operator
should consider the following possible scenarios when defining Applied Part
Types:
CONDITION 1:
If ALL of the Applied Parts are defined with the SAME Type as the
EQUIPMENT Type, then the leads will be tested TOGETHER.
CONDITION 2:
If ALL of the Applied Parts are defined with the SAME Type, but with a
different type than the EQUIPMENT Type, then the Applied Parts will be
tested INDIVIDUALLY.