Fluke Biomedical 601PRO XL User Manual
Page 38

6 0 1 P R O S E R I E S
X L
2 - 1 4
• If
YES
is selected
:
During an
Auto
sequence, when a reading is taken that
is outside the limit for the current test, the sequence will stop and a “UNIT
FAILED” message will appear with an error beep. Pressing any key will
put the operator in
Manual
mode at the test during which the failure had
occurred.
When a test is executed in an
Auto
sequence that utilizes multiple lead
types (Insulation Resistance, Patient Leakage, Patient Auxiliary Current, or
Mains on Applied Part), the 601PROXL will cycle through all of the leads
prior to determining whether a limit has been exceeded. If a failure is
detected, the operator will be placed in
Manual
mode at the failed test. The
outlet configuration will be that of the failed condition. If the test uses
applied parts, the applied part lead will be the last one tested, which is not
necessarily the lead that failed.
☛
Note:
Stop on Failure
does not apply to Accessible Voltage and
Accessible Leakage tests for IEC 1010.
• If
NO
is selected
:
The
Auto
sequence will execute until completion. The
Pass/Fail determination is performed when all tests are completed.
Default Setting:
No