X L

2 - 1 4

• If

YES

is selected

:

During an

Auto

sequence, when a reading is taken that

is outside the limit for the current test, the sequence will stop and a “UNIT

FAILED” message will appear with an error beep. Pressing any key will

put the operator in

Manual

mode at the test during which the failure had

occurred.

When a test is executed in an

Auto

sequence that utilizes multiple lead

types (Insulation Resistance, Patient Leakage, Patient Auxiliary Current, or

Mains on Applied Part), the 601PROXL will cycle through all of the leads

prior to determining whether a limit has been exceeded. If a failure is

detected, the operator will be placed in

Manual

mode at the failed test. The

outlet configuration will be that of the failed condition. If the test uses

applied parts, the applied part lead will be the last one tested, which is not

necessarily the lead that failed.

Note:

Stop on Failure

does not apply to Accessible Voltage and

Accessible Leakage tests for IEC 1010.

• If

NO

is selected

:

The

Auto

sequence will execute until completion. The

Pass/Fail determination is performed when all tests are completed.

Default Setting:

No