Computer control commands – Fluke Biomedical 601PRO XL User Manual
Page 152

6 0 1 P R O S E R I E S
X L
1 0 - 4
Computer Control Commands
ENTER DESCRIPTION
COMMAND DESCRIPTION
NOTES
[QUI]
Quit COM session
Bi-Directional
[Q]
Quit COM session
Bi-Directional
[RET]
Return Result of Last Test
[R]
Return Result of Last Test
[ACS]
Abort COM Session
Bi-Directional
[VER]
Version of 601PRO
[SN] Serial
Number
[FVER]
Returns Functionality Version
[PRT]
Print last test result to printer
[PRI]
Print Header to current printer
[PRE]
Print Header to current printer
Calibration Check Commands
COMMAND DESCRIPTION
NOTES
[CHK] Calibration
Check
Returns "PASS*" or "FAIL*" for each test’s RESISTANCE, LEAKAGE; i.e., "PASS*FAIL*" indicates RESIST
passed, LEAKAGE failed.
Outlet Control Commands
COMMAND DESCRIPTION
NOTES
[RPL]
Reverse Polarity
Turns DUT outlet ON
[NPL]
Normal Polarity
Turns DUT outlet ON
[CER] Closed
Earth
[OER] Open
Earth
[CL2] Closed
L2
[OL2] Open
L2
[OFF]
DUT Outlet OFF
[ON]
DUT Outlet ON
The DUT [ON] command always sets the outlet to Normal Polarity, Closed L2, Closed Earth. Note: The DUT
outlet polarity can be switched from Reverse Polarity to Normal Polarity with the [ON] command. To prevent
transients, the outlet control relays turns the outlet off; then, after a short delay, it turns the outlet back on with
new polarity.