6 0 1 P R O S E R I E S

X L

3 - 4

4. Selecting the Class/Type

The

CLASS/TYPE

selection sets the Type of EQUIPMENT that is being tested. This

affects the limits for the Protective Earth Resistance, Insulation Resistance L1, L2-

Case, Earth Leakage, and Enclosure Leakage Tests.

Note:

Class/Type is not used by custom standards, and hence, is not available.

The class/type should be set to the instrument's classification and type before testing.

The 601PRO defaults to the

Class/Type

setting for the most recently used test

standard.

The

Class/Type

option allows the operator to choose all valid class/type selections

for the present test standard. If the device under test has applied part leads, then they

should be entered via the

view present settings

key. The

Class/Type

option is

accessible from the

MAIN MENU

.

If the device to be tested conforms to IEC 601-1, its class/type can easily be

identified by symbols marked on the instrument.

In

Manual

mode, the Test Load used for measurements is set by the test standard

selection with the following exceptions:

⇒ The VDE Equivalent Device and VDE Equivalent Patient Leakage tests use

the IEC 601-1 Test Load exclusively.

⇒ The Accessible Voltage and Accessible Leakage tests use the IEC 1010 Test

Load exclusively. When a VDE or IEC 1010 test is selected, the Test Load is

switched to the test-specific load. When a test other than a VDE or IEC 1010

is selected, the Test Load specified by the current test standard (using the

Select Test Standard

menu) is used.