3 - 5

Determine the class/type for the instrument by using the following chart:

C l a s s / T yp e D e f i n i t i o n s

Class I

Protective Earth.
(There is no symbol
for Class I
instruments.
However, they
generally have a
protective earth
terminal.)

Class I I


Double insulation (all
plastic case)

Class IP


Internal Power Supply

Type B


Non-Isolated Applied
Part

Type BF



Isolated Applied Part

Type CF



Isolated Applied Part,
suitable for direct
cardiac application

Type F

Fixed Device

Type T

Transportable Device