Error messages, troubleshooting, and support, Error codes – Fluke Biomedical 601PRO XL User Manual
Page 161

11-1
Error Messages, Troubleshooting, and Support
The tables in this chapter list error messages that may be received during 601PRO operation,
and a brief description of the action to take.
This chapter also contains information on contacting Fluke Biomedical's Technical Service
Department for assistance and returning the 601PRO for repairs or upgrades.
1. Error Codes
2. Errors and Suggested Corrective Actions
3. Troubleshooting
4. Service
1. Error Codes
Errors that fall in the range of 0100-0900 and A100-AA00 should never be encountered.
However, if they are, you should collect as much information as possible about what you
were doing at the time of the error, determine if it is reproducible, and contact Fluke
Biomedical or your local Service Center and report the problem.
Code Description
0100 ABORT_ERR
0200
Class not valid for present standard.
0300
Type not valid for present standard and class.
0400
Standard not known.
0500
Measurement type requested invalid.
0600
No Result found for number specified.
0700
Invalid mains voltage state.
0800
Invalid applied part.
0900
Invalid test group for standard.
A100
Task control block not available.
A200
Cursor position requested invalid.
A300
A400
Failed code checksum test on power up.
A600
Flash configuration timed out.
A700
Flash read didn't match write.
A800
Device Info Record not on page boundary.
A900
RESULT structure not on page boundary.
AA00 Language files do not match size.
Chapter
11