11-1

Error Messages, Troubleshooting, and Support

The tables in this chapter list error messages that may be received during 601PRO operation,

and a brief description of the action to take.

This chapter also contains information on contacting Fluke Biomedical's Technical Service

Department for assistance and returning the 601PRO for repairs or upgrades.

1. Error Codes
2. Errors and Suggested Corrective Actions
3. Troubleshooting
4. Service

1. Error Codes

Errors that fall in the range of 0100-0900 and A100-AA00 should never be encountered.

However, if they are, you should collect as much information as possible about what you

were doing at the time of the error, determine if it is reproducible, and contact Fluke

Biomedical or your local Service Center and report the problem.

Code Description
0100 ABORT_ERR
0200

Class not valid for present standard.

0300

Type not valid for present standard and class.

0400

Standard not known.

0500

Measurement type requested invalid.

0600

No Result found for number specified.

0700

Invalid mains voltage state.

0800

Invalid applied part.

0900

Invalid test group for standard.

A100

Task control block not available.

A200

Cursor position requested invalid.

A300

not available.

A400

Failed code checksum test on power up.

A600

Flash configuration timed out.

A700

Flash read didn't match write.

A800

Device Info Record not on page boundary.

A900

RESULT structure not on page boundary.

AA00 Language files do not match size.