Fluke Biomedical 601PRO XL User Manual
Page 75

M A N U A L M O D E
3 - 3 5
To perform a VDE Equivalent Patient Leakage test:
• From the
MAIN MENU
, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Attach Patient Leads to 601PRO ECG jacks.
• From the
MAIN MENU
, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Press the
-
shortcut key
. The
Equivalent Patient Leakage
test is displayed:
baw064f.eps
• Press
START TEST (SOFT KEY 1)
to begin the test.
WARNING:
High voltage is present at the Applied Parts terminals while
measurements are being taken.
• Press the
print data
key to generate a printout of the latest measurement.
To perform a calibration from the
Equivalent Patient Leakage
test,
press
CALIBRATION (SOFT KEY 2)
.
• Follow the same process described in
Equivalent Device Leakage
.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)