M A N U A L M O D E

3 - 3 5

To perform a VDE Equivalent Patient Leakage test:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Attach Patient Leads to 601PRO ECG jacks.
• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Press the

-

shortcut key

. The

Equivalent Patient Leakage

test is displayed:

baw064f.eps

• Press

START TEST (SOFT KEY 1)

to begin the test.

WARNING:

High voltage is present at the Applied Parts terminals while

measurements are being taken.

• Press the

print data

key to generate a printout of the latest measurement.

To perform a calibration from the

Equivalent Patient Leakage

test,

press

CALIBRATION (SOFT KEY 2)

.

• Follow the same process described in

Equivalent Device Leakage

.