Test principle: vde, equivalent patient leakage – Fluke Biomedical 601PRO XL User Manual
Page 138

6 0 1 P R O S E R I E S
X L
8 - 2 4
Test Principle: VDE, Equivalent Patient Leakage
Note:
S3 = Closed
baw146f.eps
Function:
Measures the current that flows between the applied part terminals of the DUT to the PE, mains supply
conductors L1 and L2, and exposed conductive parts of the DUT in response to an isolated AC voltage
source with a magnitude of at least 110% of mains voltage.
Applicable to:
VDE 751.1