Manual mode, Connecting the device under test, Is displayed and the 601pro outlet is off.) – Fluke Biomedical 601PRO XL User Manual
Page 41: Chapter

3-1
Manual Mode
This chapter provides a brief overview of 601PRO operation and describes the power-up
sequence. It also explains how to set up the 601PRO's designation for the classification
and type of instrument to be tested, and how to run a simple test using the applicable test
standard.
1. Connecting the Device Under Test
2. The Power-Up Sequence
3. Selecting the Test Standard
4. Selecting the Class/Type
5. Saving Standard, Class, Type and Test Current
6. Using View Present Settings
7. Manual Operation
1. Connecting the Device Under Test
Before starting the test(s), ensure that the 601PRO and the device being tested are
connected properly. Refer to Figure 3-1 when connecting the Device Under Test
(DUT) to the 601PRO. Applied Parts are optional, depending on the device being
tested.
Warning:
Before operating the 601PRO, read the Warnings section on
pages vii through xi
.
When the device and the 601PRO have been connected, power up both units. (The
DUT may also be turned to ON when the
MAIN MENU
is displayed and the 601PRO
outlet is OFF.)
☛
Note:
This is good electrical practice and prevents pitting or arcing of
the 601PRO outlet or the power connector of the DUT.
Once the DUT has been plugged into the 601PRO and turned on, the operator may
freely move from test to test without changing the power switch of the DUT.
601Pr
o
XL
IN
TE
RN
AT
IO
NA
L S
AF
ET
Y A
NA
LY
ZE
R
PR
OTE
CTIV
E E
AR
TH
RE
SIS
TE
NC
E
3
PA
TIE
NT
LE
AK
AG
E u
A
6
IEC
10
10
AC
CE
SS
IBLE
V/uA
9
CU
RR
EN
T A
1
EA
RT
H
LE
AK
AG
E u
A
4
MA
INS
O
N A
P
LE
AK
AG
E u
A
7
VD
E E
QU
IVAL
EN
T
DE
VIC
E LE
AK
AG
E u
A
/
INS
UL
AT
IO
N M
2
EN
CLO
SU
RE
LE
AK
AG
E u
A
5
PA
TIEN
T A
UXIL
IAR
Y
CU
RR
EN
T uA
8
VD
E E
QU
IVA
LE
NT
PA
TIEN
T L
EA
K u
A
-
VO
LT
S
0
ES
C / S
TO
P
VIE
W P
RE
SE
NT
SE
TT
ING
S
PR
INT
HE
AD
ER
PR
INT D
ATA
To protective
earth or
enclosure
Figure 3-1: Device Under Test (DUT) Connections to the 601PRO
Chapter
3