When in the
MAIN MENU
or any manual test, the operator may also select a manual test
by pressing the
test-shortcut
keys as indicated in the following table.
Test-Shortcut
Key
Manual Test Selected
0
Mains Voltage and Dual Lead Voltage
1 Current
Consumption
2 Insulation
Resistance
3
Protective Earth Resistance
4
Earth Leakage Current
5
Enclosure Leakage Current
6
Patient Leakage Current
7
Mains on Applied Part Leakage Current
8
Patient Auxiliary Current
9
IEC 1010 Accessible Voltage and
IEC 1010 Accessible Leakage Current
/
VDE Equivalent Device Leakage Current
-
VDE Equivalent Patient Leakage Current
In addition to the above tests, Dual Lead Leakage and ECG Output can be accessed
through
the manual test menus by pressing
TESTS/AUTOMODES
, then
MANUAL
, then
*MORE
until the tests are displayed. These tests are
not
accessible
from the
test-
shortcut
keys.
Additional Features
• From the
MAIN MENU
or any manual test, press the
print header
key to print a
test header to the selected printer.
• When the
MAIN MENU
is displayed, the DUT outlet is OFF (L1, L2 and Earth
are open).
• The current standard is displayed on the
MAIN MENU
and may be changed by
selecting
SYSTEM SETUP
then
TEST STANDARD
, or by pressing the
view
present settings
key. Refer to Chapter 8, Standards and Principles, for details.
• The current class/type is displayed on the
MAIN MENU
and may be changed by
selecting
CLASS/TYPE
or by pressing the
view present settings
key. Refer to
page 3-8 in this chapter, or Chapter 2, Setting Up the 601PRO, for details.