Computer Control Commands (Cont.)

Mains on Applied Part Commands (Cont.

*

[AMAL]

Sink Current ALL-GND

AAMI Test Load

[AMRA]

Sink Current RA-GND

AAMI Test Load

[AMRL]

Sink Current RL-GND

AAMI Test Load

[AMLA]

Sink Current LA-GND

AAMI Test Load

[AMLL]

Sink Current LL-GND

AAMI Test Load

[AMV1]

Sink Current V1-V6-ALL-GND

AAMI Test Load

[ARAL]

Rev Sink Current ALL-GND

AAMI Test Load

[ARRA]

Rev Sink Current RA-GND

AAMI Test Load

[ARRL]

Rev Sink Current RL-GND

AAMI Test Load

[ARLA]

Rev Sink Current LA-GND

AAMI Test Load

[ARLL]

Rev Sink Current LL-GND

AAMI Test Load

[ARV1]

Rev Sink Current V1-V6-GND

AAMI Test Load

*

Forces L2 closed to perform test.