Fluke Biomedical 601PRO XL User Manual
Page 125

S T A N D A R D S A N D P R I N C I P L E S
8 - 1 1
IEC 1010 Controlled Power Sequence and Limits (Reflects IEC 61010-1 Second Edition, 2001-2002)
IEC 1010 STANDARD
CLASS I
CLASS II
DESCRIPTION POLARITY
CIRCUIT
Line Voltage V
N/A
N/A
NL
NL
Earth Resistance
Ω
@25A
N/A OFF 0.2 NT
Startup Delay (auto only)
Norm
Norm
Accessible Voltage RMS
Norm
Norm
33
33
Accessible Leakage RMS
µA
Norm Norm 500 500
Accessible Leakage DC
µA
Norm Norm 2000 2000
Accessible Voltage RMS
Norm
No E
55
55
Accessible Leakage RMS
µA
Norm No
E 3500 3500
Accessible Leakage DC
µA
Norm No
E 15000
15000
Power OFF Delay
Norm
No L2
Accessible Voltage RMS
Norm
No L2
55
55
Accessible Leakage RMS
µA
Norm No
L2 3500 3500
Accessible Leakage DC
µA
Norm
No L2
15000
15000
Startup Delay (auto only)
Rev
Norm
Accessible Voltage RMS
Rev
Norm
33
33
Accessible Leakage RMS
µA
Rev
Norm
500
500
Accessible Leakage DC
µA
Rev Norm 2000 2000
Accessible Voltage RMS
Rev
No E
55
55
Accessible Leakage RMS
µA
Rev No
E 3500 3500
Accessible Leakage DC
µA
Rev No
E 15000
15000
Power OFF Delay
Rev
No L2
Accessible Voltage RMS
Rev
No L2
55
55
Accessible Leakage RMS
µA
Rev No
L2 3500 3500
Accessible Leakage DC
µA
Rev No
L2 15000
15000
NOTE: The Class/Type information is not part of the IEC 1010-1 specification; hence, with the exception of Earth Resistance, the tests and
limits are the same for all Classes. The device only fails if the Accessible Voltage and Leakage are above the set limits.
Accessible Leakage RMS/DC is the same as Enclosure Leakage in IEC 601 terminology.
The limit for Protective Earth Resistance is specified as 0.1
Ω
without the mains cord impedance. Test the mains cord separately or do the
lead offset (calibration) through the mains cord. To do the specified one-minute test, Manual mode must be used.
Data is assessed based on the outlet configuration. The only requirement is for one of the three readings, Accessible Voltage, Accessible
Leakage RMS, or Accessible Leakage DC, to pass for a given outlet configuration. Alternatively stated, all three of these tests must fail for the
same outlet configuration to fail a test.