Executing auto and step mode tests – Fluke Biomedical 601PRO XL User Manual

Page 84

background image

6 0 1 P R O S E R I E S

X L

4 - 4

• Press

YES

at the following screen:

baw072f.eps

The

control number

or

template name

is displayed along with the Test Standard

and Class/Type designations that will determine the tests to perform and their limits.

When the DUT is connected and ready for testing, press

START TEST (SOFT KEY

1)

.

baw073f.eps

2. Executing Auto and Step Mode Tests

Based on the Test Standard, Class, Type and Number of Leads selected, the 601PRO

will perform the required tests in the controlled power sequence described in

Section 3: Refer to Test Limits for the Selected Standard (Chapter 8). If zero leads

are defined, no tests requiring leads will be performed. If only one lead is defined,

those tests requiring lead to lead measurements will be skipped.

Press the

esc/stop

key to abort the

Auto

or

Step

sequence at any time. Once the

esc/stop

key has been pressed, the unit may complete the test it is presently

executing before aborting.

When performing an

Auto

mode test sequence that includes Protective Earth

Resistance, the testing will pause after completion of the Protective Earth

measurement and allow the operator to adjust any connections. Press

NEXT->

to

continue the sequence (see below).

Whenever an

Auto

sequence is executed that includes a Protective Earth Resistance

Test, the operator can take as many Protective Earth Resistance readings as desired

by pressing the

START TEST

soft key. The last reading taken is the one saved. Only

the last reading taken is used for a Pass/Fail determination.