Accessible leakage – Fluke Biomedical 601PRO XL User Manual

Page 71

background image

M A N U A L M O D E

3 - 3 1

baw061f.eps

• Press

SOFT KEYS 1-3

to select the desired outlet configuration:

SOFT KEY 1

toggles the DUT outlet Polarity from Normal to DUT Off to

Reverse Polarity, or Reverse to DUT Off to Normal Polarity.

SOFT KEY 2

toggles the DUT outlet from Earth to No Earth.

SOFT KEY 3

toggles the DUT outlet from L2 to No L2.

• Modify the configuration of the front panel outlet by pressing the appropriate

SOFT KEY

on the 601PRO.

• Press the

print data

key at any time to generate a printout of the latest

measurement.

Accessible Leakage

Accessible Leakage measurements are taken from the RED input lead referenced to

the DUT Earth. Accessible Leakage cannot be selected from the

Manual

menu unless

the IEC 1010 test standard is selected.

The following outlet conditions apply when performing this test:

⇒ Normal Polarity, Earth Open, Outlet ON

⇒ Normal Polarity, Outlet ON

⇒ Normal Polarity, L2 Open, Outlet ON

⇒ Reversed Polarity, Outlet ON

⇒ Reversed Polarity, Earth Open, Outlet ON

⇒ Reversed Polarity, L2 Open, Outlet ON

To measure accessible leakage using the

test-shortcut

keys:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Press

shortcut key 9

. The

Accessible Voltage

test appears on the display.