Accessible leakage – Fluke Biomedical 601PRO XL User Manual
Page 71

M A N U A L M O D E
3 - 3 1
baw061f.eps
• Press
SOFT KEYS 1-3
to select the desired outlet configuration:
⇒
SOFT KEY 1
toggles the DUT outlet Polarity from Normal to DUT Off to
Reverse Polarity, or Reverse to DUT Off to Normal Polarity.
⇒
SOFT KEY 2
toggles the DUT outlet from Earth to No Earth.
⇒
SOFT KEY 3
toggles the DUT outlet from L2 to No L2.
• Modify the configuration of the front panel outlet by pressing the appropriate
SOFT KEY
on the 601PRO.
• Press the
print data
key at any time to generate a printout of the latest
measurement.
Accessible Leakage
Accessible Leakage measurements are taken from the RED input lead referenced to
the DUT Earth. Accessible Leakage cannot be selected from the
Manual
menu unless
the IEC 1010 test standard is selected.
The following outlet conditions apply when performing this test:
⇒ Normal Polarity, Earth Open, Outlet ON
⇒ Normal Polarity, Outlet ON
⇒ Normal Polarity, L2 Open, Outlet ON
⇒ Reversed Polarity, Outlet ON
⇒ Reversed Polarity, Earth Open, Outlet ON
⇒ Reversed Polarity, L2 Open, Outlet ON
To measure accessible leakage using the
test-shortcut
keys:
• From the
MAIN MENU
, or with the outlet unpowered, plug the DUT into the
601PRO front panel outlet, and turn on the device.
• Press
shortcut key 9
. The
Accessible Voltage
test appears on the display.