S T A N D A R D S A N D P R I N C I P L E S

8 - 3

Manual Operation

All manual tests are the same except for Leakage (referred to as Equivalent Device Leakage

Current) and Patient Leakage (referred to as Equivalent Patient Leakage Current). The following

tables detail VDE 701 and VDE 751-1 Test Limits for certain class/types.

The abbreviations are explained below:

⇒ NL: No limit. The unit will perform the test but will not apply pass/fail criteria.

⇒ NT: No test is available for this particular class/type. The unit will not take or store a

reading and will not apply pass/fail criteria. In

Manual

mode and in

Computer Control

mode, the test limit will be displayed or printed as

Inv

.

⇒ N/A: Not applicable.