beautypg.com

Ecg output – Fluke Biomedical 601PRO XL User Manual

Page 77

background image

M A N U A L M O D E

3 - 3 7

ECG Output

The 601PRO generates a series of ECG, performance, and arrhythmia waveforms to

verify the accuracy of ECG machines/monitors. ECG Output can ONLY be accessed

from the

Manual

menu.

The following outlet condition applies:

⇒ Normal Polarity, Earth Open, Outlet ON

To generate waveforms:

• From the

MAIN MENU

, or with the outlet unpowered, plug the DUT into the

601PRO front panel outlet, and turn on the device.

• Attach the device’s ECG leads to the matching 601PRO applied part terminals.
• From the

MAIN MENU

, press

TESTS/AUTOMODES

, then

MANUAL

, then

ECG OUTPUT

.

• At the

ECG/Performance Waveforms

menu, press the

SOFT KEY

below the

type of waveform/frequency to output. The outlet is powered on.

baw066f.eps

• For example, press

SOFT KEYS 1 - 3

to choose from the waveforms.

baw067f.eps

• Press

SOFT KEY 4 (

*

MORE)

for additional waveforms.

Sample waveforms are shown on the following pages.